X3 Operating Manual
Process Indicator
Sartorius
11
2.3 Display and Controls
2.3.1 Display
The display permits indication of 6-digit weight values (digit height 18 mm) with decimal point and polarity
sign.
Possible units of mass are t, kg, g or lb.
Status indication
Weight value
Mass unit
Front keys (Indicator / navigation)
Gross weight display
(G with NTEP or NSC mode)
The weight value is /- ¼ d of
zero
Net weight display
Stability of the weight value
,
Tare weight or fixtare display
Range indication
2.3.2 Front-Panel
Keys
Indicator keys
Instrument settings, set-up
Start printing
Taring, the current gross weight is stored
in the tare memory, provided that:
- weight value is stable
- indicator not in error status
(function dependent on configuration)
Set gross weight to zero, provided that:
- weight value is stable
- weight within zero setting range
(function dependent on configuration)
Selection of display mode:
gross – net – tare weight
Analog test, weighing function
Calibration and parameter input using front keys is described in chapter 4.3 .