X3 Operating Manual
Commissioning
Sartorius
69
4.5.2 Selecting the Calibration Mode
You can choose between [New] and [Modify] with the softkeys:
New Modify Param
4.5.2.1 New
Calibration
Open the menu via -[Calibration].
When you press [New], the data is set to default first and calibration is started.
SPAN and deadload
will be reset
You are prompted to confirm:
Press [Continue] for the default settings, or [Cancel] to cancel the selection.
Default settings with [New]:
WP A/Calibration
Max
3000 d
3000 kg
Scale interval
3000 d
1 kg
Deadload at
0.000000
Max at
1.000000
Not calibrated
Sensitivity
833.33
4.000000
Test
4.5.2.2 Changing the Calibration
[Modify] may be used only for minor changes (e.g. changing the dead load, adapting mV/V values
for dead load and/or Max); otherwise, always use [New].
Open the menu via -[Calibration]-[Modify].
WP A/Calibration
Max
3000 d
3000 kg
Scale interval
3000 d
1 kg
Deadload at
1.07 kg
0.000358
Max at
3000.00 kg
1.000000
Calibrated at
3000.00 kg
1.000000
Sensitivity
833.33
4.000000
by load
by mV/V
Test
For setting a new value for Dead load, press
to select [Deadload] and either enter a new
value
with [by mV/V] or discharge the scale/hopper and press [by load].
Exit calibration
without CalcTest?
When closing the menu with you are prompted whether the menu
should be closed without calculation of the test value:
Reply [Yes] to close the menu.