7.13 Counting
Purpose: Use this application to determine the number of parts of approximately equal weight.
The weight of a counted reference sample is calculated and then the objects with an
unknown piece count are weighed. The balance displays the number of parts and the
piece weight.
Minimize counting errors:
–
Ensure that the weight of individual parts is evenly distributed.
–
Larger reference piece counts lead to a higher degree of accuracy.
–
Switch on Auto Reference Updating.
t
Select the menu key in any application.
t
Select
the
Counting
symbol in the menu.
pcs
5
0.00000
Max 120 g
d=0.00001 g
Reference
y
The
Counting
application appears.
The piece count of the reference sample is displayed under
Reference
.
Reference
pcs
5
t
To set the reference sample quantity, select the gray button.
y
The
Counting
settings window appears.
t
Enter the reference sample quantity on the left.
Select a value or select
t
, enter the desired value, and confirm with
l
.
t
Select the accuracy of the weight value during counting in the middle.
y
Activate “Automatic Reference Sample Updating” on the right.
Automatic reference sample updating optimizes the counting accuracy.
The average piece weight is recalculated every time the reference sample is
updated automatically. Since the new parts broaden the basis for the calculation,
this increases the counting accuracy of the reference and with it the result.
t
Touch
l
to confirm.
t
Select
A
to zero the balance if necessary.
t
Place a container on the weighing pan.