38
System Settings (Setup Menu)
Secura, Quintix User Manual
6.2.11 More Settings
More settings
Set password
Reset settings
USB mass storage
Enable service mode
Additional service functions for the balance can be found here.
Under
Set password
(Secura
®
only):
Secura
®
models can be password-protected to prevent unauthorized personnel from
making changes to the system settings of the balance.
When a user password is set, all settings which could change the metrological
behavior of the balance are locked.
t
Select
t
.
t
Enter a number as the password.
If you have made a mistake, you can correct your entry:
–
k
button: Delete individual characters in reverse order.
–
C
button: Delete all characters.
t
Enter the same password again to ensure correct entry.
t
Confirm
with
l
.
y
The new password is only active once you have returned to the first level of the
menu.
To change the password, the old password must first be entered.
A new password can then be set.
To completely delete the password and use the balance without password protection,
leave the entry blank.
Contact the Sartorius Service Center if you forget the password.
Under
Reset settings
:
Reset all balance settings to the factory (default) settings here.
t
When the security prompt appears, select
Yes, reset
and confirm with
l
.
The balance is reset and restarts.
Under
USB mass storage
:
Use this function to register the balance memory as a USB removable data carrier on
a PC. This function is required by the Sartorius Service Center or when installing the
PC driver for the virtual COM port.
Further information can be found in Chapter “USB Port” on page 78.
Under
Enable service mode
:
This function is used by the Sartorius Service Center and can only be accessed by
authorized service technicians. The following service functions are available:
–
Preset value for minimum weighing (SQmin)
–
Enter date for next service
–
Reset all settings to factory settings
The
Calibration/Adjustment
menu also provides the following functions for
authorized Service Center personnel (depending on the model):
–
External linearization (not for use in legal metrology)
–
Set the preload
–
Delete the preload
–
Internal linearization (not for use in legal metrology)
These functions are not available for verified and sealed balances.
Visit www.scaleman.com your Authorized Sartorius Distributor