Sartorius PMA.Power PMA35001 Скачать руководство пользователя страница 9

9

Weighing Using the Recalculation Mode 

Let’s suppose that you poured in too much of one color component for a given formula (in
this example, a 4-component recipe). 
This example further assumes that you previously poured in all of the other amounts exactly
according to each of the values you entered and saved by pressing the 

L

key [MEM] (9). 

To correct for the overpour, press the 

S

key (16) to start the recalculation program; 

C

flashes

on the display. Use the 

R

key (18; up) or 

S

key (16; down) to correct the value so that it

matches the amount called for in the formula. Then press 

L

[MEM] (8) to have the scale

calculate the amount to be added for each of the components that were already poured. 
The display shows the amounts required to correct the formula up to the point at which the
overpour occurred. After the correction has been completed, you can continue filling the
remaining components.

Note:

You can correct overpours as often as needed.
Keep in mind that the total quantity of paint at the conclusion of filling increases each time
you correct a component. Press the 

c

key (7) to check how much the total quantity will be.

The 

C

stands for “Correction factor.” 

A flashing arrow 

n

on the display indicates that the weight value displayed is not valid in

legal metrology (i.e., not legal for trade).

3.

3.0

next to the weight readout.

4. Slowly pour in the first component until the 

display shows 

250 g

5. Pour in the second component until 

500 g

is 

displayed.

6. Pour in the last component 

until 1000 g

is 

displayed

This concludes the recalculation example. 
According to the display, exactly 1000 g was poured
in, but the paint can actually contains 3 kg by weight
in accordance with the factor you selected (3). 
The procedure is the same for any other conversion
factor.

Содержание PMA.Power PMA35001

Страница 1: ...98648 016 91 Installation Instructions Sartorius PMA Power Model PMA35001 Electronic Paint mixing Scales...

Страница 2: ...applications depends on the menu setting 8 L ENTER key and MEM key for paint mixing applications 9 Mounting bracket for display and control unit 10 Interface 25 contact D Sub socket 11 Grounding termi...

Страница 3: ...y designed for use in paint mixing applications This scale can be operated through the keypad in stand alone operation or using application software such as a paint mixing program from a paint manufac...

Страница 4: ...d differently The operator shall be solely responsible for any damage or injuries that occur when using cables not supplied by Sartorius Connect only Sartorius accessories and options as these are opt...

Страница 5: ...s at the place of installation To ensure that the equipment is placed in a perfectly horizontal position for consistently reproducible weighing results Always level the weighing platform again any tim...

Страница 6: ...ng or stan dard you use please contact your Sartorius office or dealer Use only original Sartorius power cables Use of cables from other manufacturers even if these units have a registered approval ra...

Страница 7: ...this example g is displayed Add the other components up to the desired weight formula Remove the filled paint can from the load plate Never use a hammer to close a paint can while is it still on the l...

Страница 8: ...of 1 0 to 5 0 or in 0 01 steps for factors of up to 1 0 Note A flashing arrow n on the display indicates that the weight value displayed is not valid in legal metrology i e not legal for trade Example...

Страница 9: ...e overpour occurred After the correction has been completed you can continue filling the remaining components Note You can correct overpours as often as needed Keep in mind that the total quantity of...

Страница 10: ...1 The readout shows the amount of the first compo nent to be added and C1 is shown in the upper right hand corner of the display 1 7 g 12 Add the first component until the readout shows 0 0 g 13 Press...

Страница 11: ...lways allow approximately 30 minutes for the scale to warm up after connecting it to AC power and before performing calibration adjustment Press the U key zero tare for 2 seconds 10000 is displayed Re...

Страница 12: ...ect the desired menu item on this level Press the L key ENTER to select the next menu level The third menu level is displayed Use the RS keys to select the desired menu item on this level Press the L...

Страница 13: ...Level 3 Level 4 SETUP Press the L key ENTER SCALE Press the L key ENTER UNIT Use the RS keys to select the desired setting e g UNIT o GRAMS Press the L key ENTER PT PD Use the RS keys to select the de...

Страница 14: ...active setting Press the c key Clear Activating the LOCK Function The LOCK function protects the scale from unauthorized use When this function is active the scale readout shows weight values only wh...

Страница 15: ...f your password in a safe place The LOCK function can be deactivated only with this password Level 1 Level 2 Level 3 INPUT Use the RS keys to select INPUT PASSWORD Press the L key ENTER Press the L ke...

Страница 16: ...ord protected Level 1 Level 2 Level 3 Level 4 SETUP Press the L key ENTER RESET Use the RS keys to select REST Press the L key ENTER MENU Use the RS keys to select MENU Press the L key ENTER YES Use t...

Страница 17: ...f changes constantly installation Too much vibration or the Access the menu to select scale is exposed to a draft the appropriate code for the ambient conditions see Oper ating Menu Configuration The...

Страница 18: ...s optimally designed to prevent damage during transport It is a good idea to save the box and all parts of the packaging for future storage or shipment of the equipment Only the original packaging pro...

Страница 19: ...e technicians on location or our Service Center in Goettingen Germany Sartorius Service Center Weender Landstrasse 94 108 37075 Goettingen Germany Prior to disposal and or scrapping of the equipment a...

Страница 20: ...for operation of the scale with a PC If you wish to create your own application software Sartorius can supply the required drivers for DOS or Windows operating systems Make sure to observe the safety...

Страница 21: ...243 132 5 Net weight approximate kg 11 4 Calibration weight kg 10 class E2 or better Power consumption VA Average 8 maximum 16 Interface port RS 232 Format 7 bit ASCII 1 start bit 1 or 2 stop bits Pa...

Страница 22: ...22...

Страница 23: ...23...

Страница 24: ...Goettingen Germany All rights reserved No part of this publication may be reprinted or translated in any form or by any means without the prior written permission of Sartorius The status of the infor...

Отзывы: