19
Error Codes
Error codes are shown on the main display for approx. 2 seconds. The program then returns
automatically to the previous mode.
Display/Problem
Cause
Solution
No segments appear
No power available
Check the power supply
on the display
The AC adapter is not plugged in
Plug in the AC adapter
Battery is drained
Replace battery; recharge battery
using external charger
oL
The load exceeds the balance
Unload the balance
capacity
UL
Weighing pan not in place
Place the weighing pan
on the balance
Something is touching
Move the object that is
the weighing pan
touching the weighing pan
disErr
Display overflow:
Reduce load on the balance
Value cannot be shown
on the display
calErr
Calibration parameter not met; e.g.:
Calibrate only when zero is displayed
– Balance not zeroed
Press (Zero) to tare the balance
– Balance is loaded
Unload the balance
appErr
Weight is too light or
Increase the weight on the balance
there is no sample on the balance
with application in use
PrtErr
Data interface for
Contact the Sartorius
printing is blocked
customer service center
bal.Err
Balance loaded or defecteive
Unload balance before switching on
when power was turned on
or contact Sartorius custumer service
sys.Err
Balance defective
Contact Sartorius custumer service
Max. weighing capacity
The balance was switched on
Place the weighing pan on the
is less than indicated
without the weighing pan in place
balance and press
e
under “Specifications”
The weight readout
The balance was not calibrated/
Calibrate/adjust the balance
is obviously wrong
adjusted before weighing
Zero the balance
Balance not zeroed
If any other errors occur, contact your local Sartorius customer service center.