15
Specific Gravity
Purpose
Use this application program to determine the specific gravity of a sample. The result is
displayed with one decimal place. Beaker and wire not included with balance.
Example:
Determine the specific gravity of a solid.
Step
Key (or instruction)
Display
1. Select application program
(Zero) > 2 sec
2. Select Specific Gravity
(F) repeatedly
3. Confirm setting
(Enter)
Symbol “ “ on the display: application is active
4. If necessary, zero the balance
(Zero)
5. Start application program
(F)
6. Confirm the display, “
airval
”
(Enter)
7. Determine weight of the sample in air:
place sample on the balance
8. Store weight-in-air value
(Enter)
9. Remove sample from balance
10. Determine weight of sample in liquid:
connect wire and set up beaker
11. Confirm the display “
Water
”
(Enter)
12. Place sample in liquid
13. Store the weight-in-liquid value and
(Enter)
view the result
14. Clear the display
(Zero)
15. Exit the Specific Gravity application
(Enter) > 2 sec
16. Reactivate Specific Gravity (if no other
(F)
application program has been selected)
17. Repeat procedure starting from Step 5.
PK Elektronik Vertriebs GmbH, E-Mail: [email protected], Internet: www.pkelektronik.com