Example:
Determining animal weight with automatic start and 20 subweighing operations
(measurements)
Parameter settings:
Applic.
:
AnimalW.
(menu code 2. 7.)
Step
Key (or instruction)
Display/Data output
1. Place animal weighing bowl
22.6
g
on the balance/scale
2. Tare the balance/scale
w
0.0
g
3. Change the number of subweighing
operations:
b
ref 30
4. Set number of measurements:
Repeatedly:
b
In increments of 1 (1, 2, 3, etc. to 100) Press briefly
In increments of 10 (10, 20, etc. to 100)
b
press
ref 20
and hold
5. Confirm number of measurements
u
+
0.0
g
*
and start automatic animal weighing
The number of measurements
remains stored in battery-backed
memory until the setting is changed
6. Place first animal in bowl.
888
The balance/scale delays the start
20
of measurements until the difference
19
between 2 measurements
....
meets the criterion
1
7. Read off the result
+
410.1
g
G
*
The result is displayed with the
“*” symbol (= calculated value)
mDef
20
and remains displayed until
x-Net +
410.1 g
the sample (animal) is removed
from the load plate (bowl)
8. Unload the balance/scale
+
0.0
g
*
9. Weigh next animal (if des.)
Next weighing series begins automatically