Totalizing
s
(Combics 2)
With this application, you can add weights to the totalizing memory. In addition to
weight values, the number of separate values added to memory is also saved
(
APPLIC.3
menu).
Features
– You can weigh up to 999 items.
– Save values automatically: Simultaneous saving of net values and calculated
values (if available).
Setting:
Applic./APPLIC.3 /TOTALIZ
menu item
3.16
.
– Save weight values and calculated values from either Application 1 (for example,
Counting, Weighing in Percent) or Application 2 (Checkweighing).
Setting:
Applic./APPLIC.3 /TOTALIZ
menu item
3.22
.
– Current transaction number displayed in the number display (indicating the
transactions already added)
– Weighing in up to a
defi
ned target, with the totalization memory c
current weight displayed in the text lines
– Save weight values manually or automatically
– Accurate calculation of total of weight values from two weighing platforms
– Activate Info mode via the
I
key
– Automatic printout when value saved
– Automatic taring of container weight.
Setting:
Applic. / aut.tare
, menu item 3.7
Exit application, delete parameters
The value of the totalizing memory remains valid until deleted by pressing the
c
key.
You can assign different functions to the
c
key for deleting applications.
When you clear applications, you can delete either the data stored for all
applications or just selected data stored for the active application.
Setting:
Applic./CLER.CF
menu item 3.24
Tare function:
If you store a tare (weight value) by pressing the
)
key, you can later enter a tare
value manually. The tare value you enter is added to the stored tare value.
Setting:
Applic. / TARE.FNC
menu item 3.25.1 (factory setting)
A tare value entered manually overwrites a stored tare value (weight value). If you
enter a tare value manually, a tare value (weight value) stored later overwrites the
manually entered value.
Setting:
Applic. / TARE.FNC
menu item 3.25.2
Restore factory default settings:
Applic./Reset
menu item 9.1.
The Combics has a totalizing memory for adding individual net and gross values.
You can save weight values in totalizing memory manually or automatically.
Setting:
Applic./APPLIC.3 /TOTALIZ
menu item
3.16.
–
Save value manually
by pressing the
O
key.
The value taken from the active platform is added to the value already saved
in totalization memory and the transaction counter value is increased by one.
When a value is added manually, the program does not check whether the
platform has been unloaded since the last time the
O
key was pressed.
Operation
Содержание Combics CAH1
Страница 125: ...Operating Instructions Combics Indicators 125 Declarations of Conformity Declarations of Conformity...
Страница 126: ...126 Operating Instructions Combics Indicators Declarations of Conformity...
Страница 127: ...Operating Instructions Combics Indicators 127 Declarations of Conformity...
Страница 128: ...128 Operating Instructions Combics Indicators EC Type approval certificate...
Страница 129: ...Operating Instructions Combics Indicators 129 certificate...
Страница 130: ...130 Operating Instructions Combics Indicators Test Certificate...
Страница 131: ...Operating Instructions Combics Indicators 131 Test Certificate...
Страница 132: ...132 Operating Instructions Combics Indicators Test Certificate...
Страница 133: ...Operating Instructions Combics Indicators 133 Plates and Markings...
Страница 134: ...134 Operating Instructions Combics Indicators Plates and Markings...
Страница 135: ...Operating Instructions Combics Indicators 135 Plates and Markings...
Страница 136: ...136 Operating Instructions Combics Indicators Plates and Markings...
Страница 137: ...Operating Instructions Combics Indicators 137 Plates and Markings...
Страница 138: ...138 Operating Instructions Combics Indicators Plates and Markings...
Страница 139: ...Operating Instructions Combics Indicators 139 Certificate...
Страница 166: ...166 Operating Instructions Combics Indicators...
Страница 167: ...Operating Instructions Combics Indicators 167...