Operating Manual BioPAT
®
Xgas Vis
Configuration
Configuration
7.4.2
Configuring Multiple Sensors with one Ethernet Modbus
Converter
If more than one BioPAT
®
Xgas Sensor is to be connected to one Perle device
(Ethernet Modbus converter), then additional configuration steps are
needed, because the default delivery Modbus address of each BioPAT
®
Xgas
Sensor is “1”.
t
Please perform the following configuration steps one after each other:
Configuration
Step
Description
1
Connect Sensor #1.
Verify that it already has Modbus address #1.
2
Label Sensor #1 with the following information:
Modbus ID : 1.
3
Disconnect Sensor #1.
4
Connect Sensor #2.
Change the Modbus address for #1 to #2. Please refer
to chapter [→7.5, page 58].
5
Label Sensor #2 with the following information:
Modbus ID : 2.
6
Disconnect Sensor #2.
7
Connect Sensor #3.
Change the Modbus address for #1 to #3.
8
Label Sensor #3 with the following information:
Modbus ID : 3.
9
Disconnect Sensor #3.
Do the same with all the others, which will be connected
to the same Perle device.