2.3
Connection of the Supplies
Required fixtures and fittings are enclosed in the delivery. Additional connection
fittings for special requirements are available; please contact your local Sartorius
Stedim Biotech representative for details.
2.3.1 Preparing the Work Place
1. During set up and installation of the bioreactor ensure that sufficient workspace
is available: (See section space requirements).
Placement of the BIOSTAT
®
CultiBag RM 600L
Connection to the facilities
Placement of the CultiBag RM 600L
Additional peripheral devices.
2. Check the mains connection:
Mains supplies in at the work place must fit to the specifications of the
bioreactor.
2.3.2 Mains
connection
CAUTION
Risk of voltage damage to equipment!
The mains supply must fulfill the specifications of the units.
Voltage fluctuations must not occur.
Verify that correct and interference free mains supply is available.
Never connect to wrong mains specifications.
Interrupt power supply before opening the cover of electrical cabinet.
Check the hardware labels to determine whether the correct mains version is supplied
and the power cords are equipped with the plugs required for the supplies.
If any of the units have incorrect mains version or power cords have wrong plugs,
please contact your local Sartorius Stedim Biotech service person.
The BIOSTAT
®
CultiBag RM 600 is verified for electrical security from plant side.
It is still necessary for customer to check the electrical security and electrical
compatibility of the internal power mains.
Delivery and Installation
11
NOTICE