If the count rate is higher than 50000 cpm, the instrument will show “OVERLOAD” and the 
dose values are not longer displayed. The dead time is approximately 35 % at this activity 
level.  The  measurement  is  still  continued  and  the  results  can  be  assessed  later  on  (after 
downloading the data to a PC).

Dead  time  ratios  higher  than  10  %  are  displayed  on  the  display  (upper  line  of  the  dose 
display)

Data transfer

The communication between the Myriam and the PC is realised by an infrared data link. The 
infrared  interface  has  to  be  connected to any USB port of 
the PC using the USB adapter cable. If the infrared interface 
is  connected  for  the  first  time,  a  driver  has  to  be  installed 
(see software manual).
The Myriam has to be placed in the middle of the interface 
unit  in  that  way  that  the  charge  receptacle  at  the 
instruments bottom panel fits into the hole of the plate. The 
instrument  has  to  be  switched  on.  Each  data  transfer  is
initiated by the PC software (remote control).
The stored time distribution, set-up parameters as well as the recent display values can be 
read even if the measurement is running. To change set-up parameters, the measurement 
must be stopped before.

Battery and Charger

The  Myriam  is  powered  by  internal  NiMH  batteries.  Using  the  “Sampling  Mode”  (running 
pump), an operation period of more than 20 hours is guaranteed. The battery recharge takes 
about two hours. A charger is included in the delivery.
If  the battery  voltage  drops  below  a  critical  level  while  the 
instrument is in the stand by mode, “BATT.  LOW” will .

A  less  or  more  filled  battery  symbol  on  the  display  shows 
the  battery  voltage  and  can  be  used  to  estimate  the 
remaining  operation  time.  Please  note  that  the  battery 
capacity depends on the temperature.
Take care for a completely charged battery at the begin of a 
longer mission.
If the voltage drops below a warn level, the sounder signalises the critical state by a short 
beep.  The  measurement  can  still  be  continued  for  a  few  hours  until  the  battery  becomes 
completely discharged. Then, the instrument switches off automatically. In this case, please 
recharge the Myriam before switching on again and downloading the data.

If the instrument is not in use for a longer period, the batteries have to be discharged and 
recharged each 2 to 3 month. The Myriam must not be charged during a sample with running 
pump.  The  charging  during  a filter analysis is possible. Please note that the charger is no 
continuous power supply. After finishing the recharge process the charger disconnects the 

Содержание MyRIAM

Страница 1: ...Myriam_EN_10 04 08 doc c SARAD GmbH User Manual MyRIAM Version 04 2008 SARAD GmbH Tel 49 0 351 6580712 Wiesbadener Stra e 10 FAX 49 0 351 6580718 D 01159 Dresden e mail support sarad de GERMANY intern...

Страница 2: ...ons 6 Gamma interference high count rates and shock rejection 6 Data transfer 7 Battery and Charger 7 Filter replacement 8 Maintenance 8 Technical Data 9 Detection thresholds and detection limits 10 S...

Страница 3: ...olled area Stopping the measurement and reading saving the acquired data Recharging the internal batteries if necessary At the end of an observation period Starting the measurement using the Filter An...

Страница 4: ...ffers a special operation mode where the pump remains switched off during the measurement The influence of the Radon daughters will become negligible and the detection limit may decreased as required...

Страница 5: ...he button for at least five Seconds If the Myriam is turned on again the data can be read via the infrared interface Pressing the button for a second time will start a new sample and the old data are...

Страница 6: ...he calculated dose value drops below the adjusted limit This situation can take place if the alert has been initiated by a high Gamma radiation or by a strong mechanical shock or continuous mechanical...

Страница 7: ...the measurement must be stopped before Battery and Charger The Myriam is powered by internal NiMH batteries Using the Sampling Mode running pump an operation period of more than 20 hours is guaranteed...

Страница 8: ...as well Maintenance The Myriam contains a high quality membrane pump to collect the aerosols on the filter This ensures a long life time and a very stable flow rate during the operation Like all move...

Страница 9: ...Integration Interval 1 255 Minutes one Minute steps Detector 150 mm ion implanted silicon detector Pump 0 25 L min Filter 0 8 m PTFE operable 1 Month at normal dust exposure Alpha Spectroscopy 2 8 10...

Страница 10: ...limits are given for the most radio toxic nuclides Pu 239 and Sr 90 without Y 90 The dose coefficients e50 of those nuclides are derived from the ICRP68 standard for inhalation 4 7E 5 Sv Bq Pu 239 an...

Страница 11: ...ion interval are stated below Fig 2 Detection Limit Threshold 3 for Ti 60 min After finishing the observation period e g one month the inhalation dose for the whole period can be determined by using t...

Страница 12: ...he first time the Windows system detects the new hardware and starts the installation assistant The installation assistant will ask for a sufficient driver which is available on the Myriam installatio...

Страница 13: ...le can be started also by pressing the button STOP Stops the running sample if not yet stopped and loads the acquired data and the instrument information from the Myriam to the PC During the data tran...

Страница 14: ...p or using the Filter Analysis mode without pump Using the radio button group Push Button locks and unlocks the push button of the Myriam The edit fields in the lower part are important for the dose c...

Страница 15: ...asurement After completion of each sequential interval a data record will be stored into the internal data logger If the check box Beep is marked the internal buzzer will sound for each keystroke Sets...

Страница 16: ...the set up dialog and the graphic window are the same like selected in the main window before independent on the recent settings of the instrument If the parameters have been transmitted to the instr...

Страница 17: ...specially at low exposures and short integration intervals the true value may be lower or higher than the displayed one It is strictly recommended to take a look at the acquired time distribution If n...

Страница 18: ...MyRIAM 18 Manual_Myriam_EN_10 04 08 doc c SARAD GmbH...

Отзывы: