MyRIAM

- 6 -

Manual_Myriam_EN_10-04-08.doc

(c) SARAD GmbH

The  physical  unit  of  the  dose  value  can  be  switched  over  by  software  between  µSv  and 
mrem.

Another  display  page  shows  the  total  alpha  and  beta  count  rate  including  the  natural 
occurring  Radon  daughter  products  in  the  unit  CPM  (counts  per  minute).  Because  Radon 
daughters are always present these values indicate a error-free operation of the instrument.

TIME: 643min
ALPHA:  3cpm
BETA: 

6cpm

If  the  count  rate  becomes  higher  than  999  /  9999  cpm  for  the  Alpha/Beta  activity,  the 
displayed  value  remains  at  999/999

9  cpm  and  the  phrase  „HIGH  COUNT“  appears  in  the 

upper display line instead of the elapsed time.

Alert situations

If the dose exceeds the user pre-set alert level, either by a 1 Minute OR an interval based 
calculation, the display automatically switches over to the related dose display page and a 
blinking symbol of a bell appears in the upper line. The LED starts to blink and the buzzer 
starts to sound for approximately 15 Seconds. The alert must be acknowledged by pressing 
the button, otherwise buzzer and LED will be activated once Minute.
If  the  alert  situation  is  still  pending  after  finishing  a  storage  interval,  a  new  alert  will  be 
generated.
The  alerts  are  re-set  if  the  calculated  dose  value  drops  below  the  adjusted  limit.  This 
situation  can  take  place  if  the  alert  has  been  initiated  by  a  high  Gamma  radiation  or by a 

strong  mechanical  shock  or  continuous  mechanical  vibrations  (see  also  chapter  “Gamma 
interference, high count rates and shock rejection”).

Gamma interference, high count rates and shock rejection

The detector of the MyRIAM is unable to distinguish between conversion electrons generated 
by Gamma radiation and electrons emitted by the collected aerosols. That means, in case of 
high Gamma exposures, a beta inhalation dose is displayed without a true contribution of the 
filter activity. In that case the dose value will decrease if the radiation field has been left. In 
case of a true inhalation dose, the activity of the LLRD at the filter remains stable.

The Alpha channel is not affected. But, it has to be attended that the detector dead time is 
increased in presence of conversion electrons  resulting in an increased Alpha detection limit 
as well.

The  detector  is  sensitive  against  strong  mechanical  shock  or  vibrations.  The  built-in 
electronic  signal  analysis  suppresses  the  noise  very  efficiently  but  continued  shock  or 
vibration results also in an increased detector dead time.

If  the  count  rate  exceeds  1000  cpm,  the  shock  rejection  procedure  is  switched  off 
automatically  to  minimise  the  dead  time.  Normally,  the  contribution  of  mechanical  shock 
becomes negligible in that case.

Содержание MyRIAM

Страница 1: ...Myriam_EN_10 04 08 doc c SARAD GmbH User Manual MyRIAM Version 04 2008 SARAD GmbH Tel 49 0 351 6580712 Wiesbadener Stra e 10 FAX 49 0 351 6580718 D 01159 Dresden e mail support sarad de GERMANY intern...

Страница 2: ...ons 6 Gamma interference high count rates and shock rejection 6 Data transfer 7 Battery and Charger 7 Filter replacement 8 Maintenance 8 Technical Data 9 Detection thresholds and detection limits 10 S...

Страница 3: ...olled area Stopping the measurement and reading saving the acquired data Recharging the internal batteries if necessary At the end of an observation period Starting the measurement using the Filter An...

Страница 4: ...ffers a special operation mode where the pump remains switched off during the measurement The influence of the Radon daughters will become negligible and the detection limit may decreased as required...

Страница 5: ...he button for at least five Seconds If the Myriam is turned on again the data can be read via the infrared interface Pressing the button for a second time will start a new sample and the old data are...

Страница 6: ...he calculated dose value drops below the adjusted limit This situation can take place if the alert has been initiated by a high Gamma radiation or by a strong mechanical shock or continuous mechanical...

Страница 7: ...the measurement must be stopped before Battery and Charger The Myriam is powered by internal NiMH batteries Using the Sampling Mode running pump an operation period of more than 20 hours is guaranteed...

Страница 8: ...as well Maintenance The Myriam contains a high quality membrane pump to collect the aerosols on the filter This ensures a long life time and a very stable flow rate during the operation Like all move...

Страница 9: ...Integration Interval 1 255 Minutes one Minute steps Detector 150 mm ion implanted silicon detector Pump 0 25 L min Filter 0 8 m PTFE operable 1 Month at normal dust exposure Alpha Spectroscopy 2 8 10...

Страница 10: ...limits are given for the most radio toxic nuclides Pu 239 and Sr 90 without Y 90 The dose coefficients e50 of those nuclides are derived from the ICRP68 standard for inhalation 4 7E 5 Sv Bq Pu 239 an...

Страница 11: ...ion interval are stated below Fig 2 Detection Limit Threshold 3 for Ti 60 min After finishing the observation period e g one month the inhalation dose for the whole period can be determined by using t...

Страница 12: ...he first time the Windows system detects the new hardware and starts the installation assistant The installation assistant will ask for a sufficient driver which is available on the Myriam installatio...

Страница 13: ...le can be started also by pressing the button STOP Stops the running sample if not yet stopped and loads the acquired data and the instrument information from the Myriam to the PC During the data tran...

Страница 14: ...p or using the Filter Analysis mode without pump Using the radio button group Push Button locks and unlocks the push button of the Myriam The edit fields in the lower part are important for the dose c...

Страница 15: ...asurement After completion of each sequential interval a data record will be stored into the internal data logger If the check box Beep is marked the internal buzzer will sound for each keystroke Sets...

Страница 16: ...the set up dialog and the graphic window are the same like selected in the main window before independent on the recent settings of the instrument If the parameters have been transmitted to the instr...

Страница 17: ...specially at low exposures and short integration intervals the true value may be lower or higher than the displayed one It is strictly recommended to take a look at the acquired time distribution If n...

Страница 18: ...MyRIAM 18 Manual_Myriam_EN_10 04 08 doc c SARAD GmbH...

Отзывы: