User Manual & Datasheet V 2.3

OPEN

ENTER

COVER

UNCOVER

SENSITIVITY

FAILSAFE

VITAL

STATUS

LED

RED

GREEN

Figure 17: Failsafe Low

2. When not in contact with the material, LED turns RED.

14

Troubleshooting & Indications

14.1

Output Indications

Green LED Glows When:

Fork is Uncovered and Failsafe is High

Fork is Covered and Failsafe is Low

Red LED Glows When:

Fork is Covered and Failsafe is High

Fork is Uncovered and Failsafe is Low

14.2

Troubleshooting

The instrument is working but there may be solid par-
ticles in the application material that may be clogging
the fork tines. Make sure particle size is 12 mm.

Heavy build-up of application material can dampen the
fork oscillations. In this case, the fork requires to be
cleaned. To avoid build up change sensitivity value to
1.

If the vibrating fork is not sensing application media,
consider changing the sensitivity.

If instrument is not powering on (No LED is glowing).
Check input supply voltage. It should be in the range
mentioned in Electrical Specifications

Connectivity between Vibating fork and Electron-
ics is broken

In failsafe ’High’ mode, instrument will switch to
ALARM

In failsafe ’Low’ mode, instrument will switch to
NORMAL

15

Maintenance

The electronics of Vital instrument needs no mainte-

nance. When cleaning and checking the vessel, free the
tuning fork from deposits. If the material has tendency to
form a hard sticky deposit, the instrument must be checked
more often. Make sure that the cable ducts and the lid are
tightly sealed so that no moisture seeps into the instrument.

16

Customer Support

Thank you for going through the instructions given in this

manual. To further ease the process of installation and use,
we have developed special demo videos which are hosted
on YouTube.

Sapcon’s YouTube channel, SAPCON INSTRUMENTS,
lists all these videos: https://goo.gl/dnxfcz

Should you require further information regarding installa-

tion, use or working of the instrument, please don’t hesitate
to contact us. Kindly provide the following information at the
time of contacting:

Instrument Model and Serial Number

Purchase Order Number and Date of Purchase

Description of the query

Your contact details

In an attempt to serve you better, we are open seven days
a week (9:30am to 7:30pm). We are available at:

www.sapconinstruments.com

[email protected]

+91-731-4757575

Sapcon Instruments Pvt.Ltd.

R

11

Содержание Vital

Страница 1: ...VITAL Compact Vibrating Fork Level Switch INSTRUCTION MANUAL Version 2 3 ...

Страница 2: ...ations 6 7 Mechanical Specifications 7 8 Installation Handling Guidelines 7 9 Electrical Connections 8 10 Sensitivity Settings 9 11 Cover Delay 9 12 Uncover Delay 10 13 Failsafe Settings 10 14 Troubleshooting Indications 11 14 1 Output Indications 11 14 2 Troubleshooting 11 15 Maintenance 11 16 Customer Support 11 17 Product Selection Order Code 12 Sapcon Instruments Pvt Ltd R 2 ...

Страница 3: ...on 9 8 Setting Sensitivity Value 9 9 Saving Sensitivity Value 9 10 Cover Delay Switch Position 9 11 Setting Cover Delay 10 12 Saving Cover Delay 10 13 Uncover Delay Switch Position 10 14 Setting Uncover Delay 10 15 Saving Uncover Delay 10 16 Failsafe High 10 17 Failsafe Low 11 List of Tables 1 Electrical Specifications 6 2 Application Specifications 6 3 Mechanical Specifications 7 Sapcon Instrumen...

Страница 4: ...ual instrument housing in terms of dimensions color and design Please refer to GA drawings for dimensional details Values of performance described in this manual were obtained under ideal testing conditions Hence they may differ under industrial environment and settings General Instructions Instrument shouldn t block the material filling inlet Secure the cover of housing tightly Tighten the cable ...

Страница 5: ...tude is detected and used for switch ing the output The output contacts of the relay are in turn used for annunciation and or control Figure 2 Description of Parts 3 Features Low Power Consumption Less heat Long life Fast switching minimum of 0 5s availbale on request with FS option Self diagnosis External LED indication available with WL option Independent of material s electrical properties Comp...

Страница 6: ...e High For High Level Close Fail safe Low For Low Level Time Delay Settings 1 25 seconds For both Covered and Uncovered Delays Operating Temperature Ambient Temperature 0 C to 65 C Process Temperature 0 C to 100 C with option HT Relay Rating 6 Amp at 230 VAC Table 1 Electrical Specifications 6 Application Specifications Please refer to Table 2 for Application Specifications PARAMETER VALUE Sensiti...

Страница 7: ...inlet source a canopy called baffle of appropriate strength and size should be welded right above the fork as shown While installing the instrument please take care of the following points The instrument shouldn t block the material filling inlet Secure the cover of housing tightly Tighten the cable glands For side mounting provide a baffle to prevent the application material from falling on the f...

Страница 8: ...ween 60 C and 80 C While screwing the instrument the hexagonal mounting bush should be turned and not the housing 9 Electrical Connections Please refer to Figure 5 for electrical connections OPEN ENTER COVER UNCOVER SENSITIVITY FAILSAFE VITAL STATUS LED RED GREEN L N E Relay 1 PNP 1 2 3 4 5 6 7 8 9 NC C NO OUT AC Supply DC Supply L Live N Neutral 90 265 V AC 18 55 V DC RL Available only in DC Powe...

Страница 9: ...gure 8 Setting Sensitivity Value 4 The value of sensitivity is now entered but not saved To save and test the sensitivity set the SENSITIVITY switch back to OPEN position as shown in Figure 9 The STATUS LED will come back to its original posi tion OPEN ENTER COVER UNCOVER SENSITIVITY FAILSAFE VITAL STATUS LED RED GREEN Figure 9 Saving Sensitivity Value 11 Cover Delay When the application material ...

Страница 10: ... in Figure 14 The STATUS RED LED will start blinking Count the number of blinks After setting the value release the ENTER key OPEN ENTER COVER UNCOVER SENSITIVITY FAILSAFE VITAL STATUS LED RED GREEN Figure 14 Setting Uncover Delay 4 Uncover Delay is entered but not saved To save and test the Uncover Delay set the UNCOVER switch back to OPEN position as shown in figure 15 The STATUS LED will come b...

Страница 11: ... failsafe High mode instrument will switch to ALARM In failsafe Low mode instrument will switch to NORMAL 15 Maintenance The electronics of Vital instrument needs no mainte nance When cleaning and checking the vessel free the tuning fork from deposits If the material has tendency to form a hard sticky deposit the instrument must be checked more often Make sure that the cable ducts and the lid are ...

Страница 12: ...Depends on Vibrating Fork FS Time Delay Adjustable from 0 5 to 20 seconds for probe covered or uncovered Only with VF105S VF130 NS Normal Time Delay Adjustable from 2 to 20 seconds for probe covered or uncovered Mounting MB15MS Screwed Thread BSP 1 1 2 MS Plated MB15S4 Screwed Thread BSP 1 1 2 SS 304 MB15S6 Screwed Thread BSP 1 1 2 SS 316 MN15MS Screwed Thread NPT 1 1 2 MS Plated MN15S4 Screwed Th...

Страница 13: ... 125mm Material SS 316 Only with 1 55H VF185 Total Length 185mm Tines 155mm Material SS 316 Only with 2H 2H30H 2 5H30H Finish HB Standard FB Fully Buffed Extension Material Depends on Probe Length Only with 2H30H 2 5H30H EGI GI Galvanized Iron ES4 SS 304 ES6 SS 316 Standoff Material Depends on Operating Temperature Only with 20T STGI GI Galvanized Iron STS4 SS 304 STS6 SS 316 Operating Temperature...

Отзывы: