User Manual & Datasheet V 2.3

OPEN

ENTER

COVER

UNCOVER

SENSITIVITY

FAILSAFE

VITAL

STATUS

LED

RED

GREEN

Figure 11: Setting Cover Delay

OPEN

ENTER

COVER

UNCOVER

SENSITIVITY

FAILSAFE

VITAL

STATUS

LED

RED

GREEN

Figure 12: Saving Cover Delay

The STATUS LED will start blinking RED if the switch

point is reached. It will blink for the number of seconds for
which the cover delay is set. 1 blink is equal to 1 second
during switching. A maximum of 25 seconds can be set.

12

Uncover Delay

When the application material uncovers vital’s fork tines,

the changeover of the output can be delayed by a pre-
determined time. This time is called UNCOVER Delay. For
a different value of Uncover Delay, the number of blinks can
be adjusted as per requirement.

Note:

You can set the value of UNCOVER DELAY between
1-25 secs.

Follow the below procedure for setting Uncover

Delay

1. Ensure that all DIP switches are in OPEN position as

shown in Figure 6. Make sure that STATUS LED is not
blinking for Error.

2. To set the Uncover Delay, set the UNCOVER switch to

CLOSE position as shown in Figure 13.
(CLOSE is the opposite of OPEN for a DIP switch.)
The STATUS RED LED will glow.

OPEN

ENTER

COVER

UNCOVER

SENSITIVITY

FAILSAFE

VITAL

STATUS

LED

RED

GREEN

Figure 13: Uncover Delay Switch Position

3. Press ENTER and keep it pressed as shown in Figure

14. The STATUS RED LED will start blinking. Count
the number of blinks. After setting the value release
the ENTER key.

OPEN

ENTER

COVER

UNCOVER

SENSITIVITY

FAILSAFE

VITAL

STATUS

LED

RED

GREEN

Figure 14: Setting Uncover Delay

4. Uncover Delay is entered, but not saved. To save and

test the Uncover Delay, set the UNCOVER switch back
to OPEN position as shown in figure 15. The STATUS
LED will come back to its original position.

OPEN

ENTER

COVER

UNCOVER

SENSITIVITY

FAILSAFE

VITAL

STATUS

LED

RED

GREEN

Figure 15: Saving Uncover Delay

5. To test, dip vital into the application material until the

switching point is achieved.

6. The STATUS LED will start blinking GREEN if the

switch point is achieved. It will blink for the number
of seconds for which the Uncover Delay is set.

13

Failsafe Settings

In a condition of device failure, known errors and input

power failure the outputs of the device resemble the ALARM
condition. This is meant to prevent overflow or dry run con-
ditions in case of failures.

Prevent Overflow - High Level Switch

Failsafe High (de-

fault) is set by moving the Failsafe switch to OPEN position.

1. When not in contact with the material, LED turns

GREEN.

2. When in contact with the material, LED turns RED.

OPEN

ENTER

COVER

UNCOVER

SENSITIVITY

FAILSAFE

VITAL

STATUS

LED

RED

GREEN

Figure 16: Failsafe High

Prevent Dry run - Low Level Switch

Failsafe Low is set

by moving the Failsafe switch to CLOSE position

1. When in contact with the material, LED turns GREEN.

Sapcon Instruments Pvt.Ltd.

R

10

Содержание Vital

Страница 1: ...VITAL Compact Vibrating Fork Level Switch INSTRUCTION MANUAL Version 2 3 ...

Страница 2: ...ations 6 7 Mechanical Specifications 7 8 Installation Handling Guidelines 7 9 Electrical Connections 8 10 Sensitivity Settings 9 11 Cover Delay 9 12 Uncover Delay 10 13 Failsafe Settings 10 14 Troubleshooting Indications 11 14 1 Output Indications 11 14 2 Troubleshooting 11 15 Maintenance 11 16 Customer Support 11 17 Product Selection Order Code 12 Sapcon Instruments Pvt Ltd R 2 ...

Страница 3: ...on 9 8 Setting Sensitivity Value 9 9 Saving Sensitivity Value 9 10 Cover Delay Switch Position 9 11 Setting Cover Delay 10 12 Saving Cover Delay 10 13 Uncover Delay Switch Position 10 14 Setting Uncover Delay 10 15 Saving Uncover Delay 10 16 Failsafe High 10 17 Failsafe Low 11 List of Tables 1 Electrical Specifications 6 2 Application Specifications 6 3 Mechanical Specifications 7 Sapcon Instrumen...

Страница 4: ...ual instrument housing in terms of dimensions color and design Please refer to GA drawings for dimensional details Values of performance described in this manual were obtained under ideal testing conditions Hence they may differ under industrial environment and settings General Instructions Instrument shouldn t block the material filling inlet Secure the cover of housing tightly Tighten the cable ...

Страница 5: ...tude is detected and used for switch ing the output The output contacts of the relay are in turn used for annunciation and or control Figure 2 Description of Parts 3 Features Low Power Consumption Less heat Long life Fast switching minimum of 0 5s availbale on request with FS option Self diagnosis External LED indication available with WL option Independent of material s electrical properties Comp...

Страница 6: ...e High For High Level Close Fail safe Low For Low Level Time Delay Settings 1 25 seconds For both Covered and Uncovered Delays Operating Temperature Ambient Temperature 0 C to 65 C Process Temperature 0 C to 100 C with option HT Relay Rating 6 Amp at 230 VAC Table 1 Electrical Specifications 6 Application Specifications Please refer to Table 2 for Application Specifications PARAMETER VALUE Sensiti...

Страница 7: ...inlet source a canopy called baffle of appropriate strength and size should be welded right above the fork as shown While installing the instrument please take care of the following points The instrument shouldn t block the material filling inlet Secure the cover of housing tightly Tighten the cable glands For side mounting provide a baffle to prevent the application material from falling on the f...

Страница 8: ...ween 60 C and 80 C While screwing the instrument the hexagonal mounting bush should be turned and not the housing 9 Electrical Connections Please refer to Figure 5 for electrical connections OPEN ENTER COVER UNCOVER SENSITIVITY FAILSAFE VITAL STATUS LED RED GREEN L N E Relay 1 PNP 1 2 3 4 5 6 7 8 9 NC C NO OUT AC Supply DC Supply L Live N Neutral 90 265 V AC 18 55 V DC RL Available only in DC Powe...

Страница 9: ...gure 8 Setting Sensitivity Value 4 The value of sensitivity is now entered but not saved To save and test the sensitivity set the SENSITIVITY switch back to OPEN position as shown in Figure 9 The STATUS LED will come back to its original posi tion OPEN ENTER COVER UNCOVER SENSITIVITY FAILSAFE VITAL STATUS LED RED GREEN Figure 9 Saving Sensitivity Value 11 Cover Delay When the application material ...

Страница 10: ... in Figure 14 The STATUS RED LED will start blinking Count the number of blinks After setting the value release the ENTER key OPEN ENTER COVER UNCOVER SENSITIVITY FAILSAFE VITAL STATUS LED RED GREEN Figure 14 Setting Uncover Delay 4 Uncover Delay is entered but not saved To save and test the Uncover Delay set the UNCOVER switch back to OPEN position as shown in figure 15 The STATUS LED will come b...

Страница 11: ... failsafe High mode instrument will switch to ALARM In failsafe Low mode instrument will switch to NORMAL 15 Maintenance The electronics of Vital instrument needs no mainte nance When cleaning and checking the vessel free the tuning fork from deposits If the material has tendency to form a hard sticky deposit the instrument must be checked more often Make sure that the cable ducts and the lid are ...

Страница 12: ...Depends on Vibrating Fork FS Time Delay Adjustable from 0 5 to 20 seconds for probe covered or uncovered Only with VF105S VF130 NS Normal Time Delay Adjustable from 2 to 20 seconds for probe covered or uncovered Mounting MB15MS Screwed Thread BSP 1 1 2 MS Plated MB15S4 Screwed Thread BSP 1 1 2 SS 304 MB15S6 Screwed Thread BSP 1 1 2 SS 316 MN15MS Screwed Thread NPT 1 1 2 MS Plated MN15S4 Screwed Th...

Страница 13: ... 125mm Material SS 316 Only with 1 55H VF185 Total Length 185mm Tines 155mm Material SS 316 Only with 2H 2H30H 2 5H30H Finish HB Standard FB Fully Buffed Extension Material Depends on Probe Length Only with 2H30H 2 5H30H EGI GI Galvanized Iron ES4 SS 304 ES6 SS 316 Standoff Material Depends on Operating Temperature Only with 20T STGI GI Galvanized Iron STS4 SS 304 STS6 SS 316 Operating Temperature...

Отзывы: