28
RECYCLE OF NICKEL-CADMIUM BATTERY
MDF-U3086S
The unit contains a rechargeable battery. The battery is recyclable. At the end
of it’s useful life, check with you local solid officials option or proper disposal.
This product has an electric battery.
Please recycle it for effective utilization of resources and protection of our
environment.
Thank you for your cooperation.
Label indication is obliged to comply with Taiwanese battery regulation.
1. Location of a nickel-cadmium battery
This unit is provided a nickel-cadmium battery for the power failure and temperature warning device.
The battery is located in the control box behind the control panel.
2. Removal of nickel-cadmium battery
1) Turn off the power switch and battery switch, and disconnect the power supply plug.
2) As shown in the Fig. 1, remove 2 screws on the right side, fixing the grille by using a screw driver.
3) Remove 4 screws fixing the control panel cover to take off the cover. (Fig. 2)
4) Remove 4 screws fixing the control panel and separate the panel from the frame. (Fig. 3)
5) Remove the control box cover behind the control panel. (Fig. 4)
6) Take out the battery by removing the clip fixing the battery.
3. Handling of battery
Cover the battery terminal with an insulating tape to avoid the short circuit. Then follow the procedure
for recycling or proper disposal.
廢電池
請回收
Fig. 2
Screw
Screw
Fig. 3
Fig. 4
Fig. 1
Grille
Screw
Screw
Screw
Battery
Clip