71
25
Sensor box heater
abnormality
If the sensor box heater goes open
circuit, or the sensor box relay short
circuit.
E15 is displayed
alternately
with the temperature on
the
temperature indicator.
Intermittent tone
Remote alarm
Disconnection of
sensor for each
heater
If the relay of main heater, bottom
heater or sensor box heater goes open
circuit.
E16 is displayed
alternately
with the temperature on
the
temperature indicator.
Intermittent tone
Remote alarm
Air pump (for
sample
air or auto-zero)
failure
If the air pump (sampling or auto zero)
does not operate, or if there is
something wrong in the gas piping.
E17 is displayed
alternately
with the temperature on
the
temperature indicator.
Intermittent tone
CO
2
valve close.
Remote alarm
Low humidifying
water
If the water in the pan is about 1 liter.
RH PAN lamp blinks.
-----
-----
UV lamp failure
[When MCO-18UVS is installed]
When the UV lamp is burned out.
E18 is displayed
alternately
with the temperature on
the
temperature indicator.
Intermittent tone
Remote alarm
Recommendation of
new UV lamp
[When MCO-18UVS is installed]
The accumulated ON time is over
about
1,000 hrs.
UV indicator blinks when
UV lamp is OFF.
----- -----
• The alarm can be canceled by pressing the BUZZER key, but the remote alarm cannot be silenced. And the upper
limit temperature alarm cannot be silenced with the BUZZER key.
• E01 is cleared automatically when the gas is connected correctly and the buzzer is silenced with the BUZZER key.
When MCO-21GC is installed, press BUZZER key to silence the alarm after changeover of gas supply line.
• If one of E05 to E17 (except for E12, E13, and E14) is displayed, consult with a Sanyo sales representative or
agent.
SETTING OF ALARM RESUME TIME
The alarm buzzer is silenced by pressing BUZZER key on the control panel during alarm condition.
The buzzer will be activated again after certain suspension if the alarm condition is continued. The
suspension time can be set by following the procedure shown in the table below.
The example in the table is based on the assumption that the desired duration is 20 minutes.
Note:
The duration is set in 30 minutes at the factory.
Table Changing procedure for alarm resume time (Ex: change from 30 minutes to 20 minutes)
Description of operation
Key operated
Indication after operation
1
----
The current chamber temperature is
displayed.
2 Press CAL key for 5 seconds.
CAL
The left digit is flashed.
The settable digit is shifted.
3 Set the figure to F25 with the digit
shift key and numerical value shift
key.
When pressed, the figure of settable
digit changes.
4 Press ENT key.
ENT
The current setting is displayed.
The middle digit is flashed.
5
Set the figure to 020 with the
numerical value shift key.
When pressed, the figure of settable
digit changes.
6 Press ENT key.
ENT
The setting is memorized and
the current chamber temperature is
displayed.
• The settable alarm resume time are 0, 10, 20, 30, 40, 50, or 60 minutes (The setting is 000, 010, 020,
030, 040, or 060 respectively). The buzzer would not reset if the reset time is set in 000.
• The set mode returns to the temperature display mode automatically when 90 seconds has passed
without any key operation. In this case, any setting before pressing ENT key is not memorized.
Содержание MCO-18AIC
Страница 1: ...Service Manual CO2 Incubator MCO 18AIC FILE No SM9910023 ...
Страница 5: ......
Страница 8: ...5 Dimensions ...
Страница 9: ...6 ...
Страница 10: ......
Страница 23: ...20 Wiring Diagram ...
Страница 24: ...21 Circuit Diagram ...
Страница 25: ...22 Components on PCB ...
Страница 29: ......
Страница 49: ...46 ...
Страница 73: ......
Страница 79: ......
Страница 86: ......
Страница 87: ...JK1148 250B ...