2
gas density calibration and Auto Zero adjustment
MCO-18AIC provides both the manual CO
2
gas density calibration and the automatic Auto
Zero adjustment.
1. CO
2
gas density calibration
When either the CO
2
density SV or the display value is not match with the actual value,
calibrate CO
2
gas density according with “CO
2
calibration” (Page 9) to input the actual
value again.
2. Auto Zero adjustment
Refer to “Auto Zero adjustment”
3. The relation the CO
2
gas density calibration with the Auto Zero adjustment
The description is as follow.
(
Ex: CO
2
density = 5%
)
Point calibrated by auto zero adjustment
Set density(%CO
2
)
0
0
5
A
ctual density
(%C
O
2
)
5
Point calibrated by CO
2
gas density
calibration
Ideal line for the set value and the actual
value.
(Note) In fact, it could not make a straight
line due to the characteristics of the
sensor.