OPERATING INSTRUCTIONS
Change of setting for UV lamp ON period
The clauses below are applicable when an optional UV system kit MCO-18UVS is installed.
Follow the procedure below when changing the setting for UV lamp ON period.
Basic operation sequence (Example: change of UV lamp ON period from 5 minutes to 3 minutes)
Description of operation
Key operated
Indication after operation
1 Press CAL key for 5 seconds.
CAL
The left digit in the temperature
indicator is flashed.
When pressed, the changeable digit is shifted.
2
By pressing digit shift key and
numerical value shift key, set the
figure to F01.
When pressed, the figure of settable digit
changes.
3 Press ENT key.
ENT
The current setting is displayed
in the CO
2
density indicator.
When pressed, the changeable digit is shifted.
4
By pressing digit shift key and
numerical value shift key, set the
figure to 003.
When pressed, the figure of settable digit
changes.
5 Press ENT key.
ENT
Set value is memorized and the display return to
normal display mode.
• The available set range for the UV lamp is between 0 and 30 minutes (000 to 030). When set to 000,
the UV lamp is not turned on.
• The UV lamp is turned off during ON period when the outer door is opened. After closing the outer
door, the lamp turns on during predetermined period.
• Condensation will occur and/or temperature distribution may be affected due to the heat of the UV lamp
when the setting of the lamp operation is longer than 5 minutes or if only the outer door is opened
repeatedly.
• For the replacement of UV lamp, contact Sanyo sales representative or gent.
Pressing CAL key for about 5 seconds leads the calibration mode. In the calibration mode, the
calibration of temperature and CO
2
density is possible. Wrong key operation affects the basic
performance. Never touch any other keys on the control panel in the event of pressing CAL key
accidentally. After about 90 seconds, the unit returns to chamber temperature display mode
automatically.
Содержание MCO-18AIC
Страница 1: ...Service Manual CO2 Incubator MCO 18AIC FILE No SM9910023 ...
Страница 5: ......
Страница 8: ...5 Dimensions ...
Страница 9: ...6 ...
Страница 10: ......
Страница 23: ...20 Wiring Diagram ...
Страница 24: ...21 Circuit Diagram ...
Страница 25: ...22 Components on PCB ...
Страница 29: ......
Страница 49: ...46 ...
Страница 73: ......
Страница 79: ......
Страница 86: ......
Страница 87: ...JK1148 250B ...