Followings are initial values in non-volatile memory:
Internal temperature SV
: 37.0
°
C
CO
2
level SV
: 0.0%
Balance value of humidifying heater (F06)
: 12
Balance value of door heater (F07)
: 4
CO
2
zero adjustment data (F10)
: 5000
CO
2
span data (F11)
: 5000
Internal temperature zero adjustment data (F15)
: 008
Key Lock data
: Key Lock OFF
Auto Changer data (the 3
rd
digit in F08)
: Auto Changer OFF
Demo Mode data (the 2
nd
digit in F08)
: Demo Mode OFF
Auto Zero (the 1
st
digit in F08)
: Auto Zero ON
Input “F13” on temperature display and press ENT key to display “000” on
CO
2
display, then change the 1
st
digit to “2” and press ENT key to initialize
CO
2
zero adjustment data and CO
2
span data in non-volatile memory.
Note) Service code should not be initialized.
F14: Check the performance for LED, drivers
<Direction
for
use>
Input “F14” on temperature display and press ENT key to perform with lamp
and the all digits flashing at every 0.5seconds intervals. Press ENT key to
stop performance.
If there is no key operation for approx. 90seconds, automatically comes to
end
F15:
Display an adjustment value for temperature zero point (initial: 0.8
°
C)
<Direction
for
use>
Input “F15” on temperature display and press ENT key to display an
adjustment value on CO
2
display for temperature zero point. If you press ENT
key or leave the device for 90seconds, automatically returns to PV display.
(Ex.) In “37.0” is displayed on the control panel and actual temperature is “37.8”, if
you perform temperature calibration (actual temperature “37.8” is input),
“+0.8” would be housed in F15. In Function Mode, input F15 and press ENT
key to display the value on CO
2
display.
F21: Set serial communication ID (initial: 000)
<Direction
for
use>
Input “F21” on temperature display and press ENT key to show “000” on CO
2
display. Then input the value that identifies the product (serial communication
ID Number) and press ENT key to set serial communication ID. The range is
000~255. (000: communication OFF)
F22: Set serial communication mode (initial: 000)
<Direction
for
use>
Input “F22” on temperature display and press ENT key to show “000” on CO
2
display. Then change the 3
rd
digit to set the communication mode.
Control mode (the 3
rd
digit) 0: Local (initial)
1: Remote
Baud rate (the 2
nd
digit) 0: 2400bps
1: 4800bps
2: 9600bps
F24: Interlock the remote alarm with buzzer (initial: 000)
<Direction
for
use>
Input “F24” on temperature display and press ENT key to show “000/001” on
CO
2
display. Then change the 1
st
digit to “000” to set the interlock mode.
000: Buzzer/Remote alarm not interlocked
001: Buzzer/Remote alarm interlocked
Содержание MCO-18AIC
Страница 1: ...Service Manual CO2 Incubator MCO 18AIC FILE No SM9910023 ...
Страница 5: ......
Страница 8: ...5 Dimensions ...
Страница 9: ...6 ...
Страница 10: ......
Страница 23: ...20 Wiring Diagram ...
Страница 24: ...21 Circuit Diagram ...
Страница 25: ...22 Components on PCB ...
Страница 29: ......
Страница 49: ...46 ...
Страница 73: ......
Страница 79: ......
Страница 86: ......
Страница 87: ...JK1148 250B ...