[8] SPECIFICATIONS
8-1 General Specifications
Operation
method
Delta-sigma modulation
LCD display
Main display
50,000 counts: DCV, DCmV, ACV, ACmV,
DCA, DCmA, DC
μA,
ACA, ACmA, AC
μA,
Resistance, Continuity
500,000 counts: DCV, DCmV
99,999 counts: Logic-Level Frequency (Hz)
20,000 counts: Diode
12,500 counts: Temperature (
℃
)
22,820 counts: Temperature(
°
F
)
9,999 counts: nS, dBm, Duty Cycle
5,000 counts: Capacitance
Bar graph: Up to 41 segments
Sub display
5,000 counts: ACV, ACmV,
ACA, ACmA, AC
μA
9,999 counts: Frequency (Hz),
Loop-current % 4
~
20 mA
Over-range
indication
Over-range input turns on "OL" indicator at the numeric part.
Sampling rate
Numeric part 5 times / sec.(50,000-count mode)
1.25 times/sec. (500,000-count mode)
Bar graph part 60 times / sec.
Low battery
indication
Decreasing the internal battery voltage to approx.
7 V turns the battery mark on.
Operating
conditions
Altitude: < 2,000 m Pollution degree: II
Operating
temperature/
humidity
5
℃
to 40
℃
: non-condensing
5
℃
to 31
℃
: 80 %RH (Max.)
31
℃
to 40
℃
: decreasing 80 % to 50 % linearly
Storage
temperature/
humidity
-10
℃
to 40
℃
: 80 %RH (Max.) non-condensing
(with battery removed)
40
℃
to 50
℃
: 70 %RH (Max.) non-condensing
(Remove the battery, if the equipment
is not going to be used for a long time.)
Temperature
coefficient
0.15 x (accuracy @23±5
℃
)/
℃
@(0
℃
to 18
℃
or
28
℃
to 40
℃
)
Power source Single alkaline 9 V battery 6LR61(IEC6LF22, NEDA1604A)
AC sensing
method
True RMS (AC coupling)
OVERVOLTAGE CATEGORY
Equipment of CAT II: Primary cable runs of power-consuming
equipments from a wall socket.
Equipment of CAT III: Primary cable runs of equipments directly
connected to a distribution board and cable
runs from a distribution board to wall sockets.
Equipment of CAT IV: Cable runs from an incoming line to a
distribution board.
Auto Power
Saving
Approx. 17 minutes after the last operation
Safety
Compliances
IEC61010-1, IEC61010-2-030, IEC61010-2-033
IEC61010-031
Category II for 1000 V ac and dc
Category III for 600 V ac and dc
EMC
Meets EN61326-1:2006
In an RF field of 3 V/m:
Capacitance function is not specified
Other function ranges:
Total Accuracy=
±
(
Specified % rdg + 1000 digits)
Performance above 3 V/m is not specified
Dimensions
without holster Approx.
L 175 mm×W 80 mm×H 40 mm
with holster Approx.
L 184 mm×W 86 mm×H 52 mm
Mass
without holster Approx. 360 g
with holster Approx. 430 g
Power
consumption
Approx. 58 mW / Approx. 72 mW with LPF in use
Approx. 0.63 mW (Auto Power Saving)
Battery life
Approx. 100 hours (DCV measurement)
Accessories
Test leads (TL-23a),
Holster (H-700) with light shieliding magnet cap,
K-type thermocouple (K-250PC),
Instruction manual
Содержание PC7000
Страница 1: ...PC7000 DIGITAL MULTIMETER 取扱説明書 INSTRUCTION MANUAL ...
Страница 17: ... 30 29 備 考 m m m m D 時にはバーグ ラフは表示されません m m m m D ...
Страница 25: ... 46 45 備考 4 20 mA 測定では DC 4 mA を 0 とし DC 20 mA を 100 とした場合のパーセンテージ を表示します m 4 20 mA m m m m m Hz ...
Страница 40: ...PC7000 DIGITAL MULTIMETER INSTRUCTION MANUAL ...
Страница 56: ... 30 29 Note The display style of m m m m or D does not show the bar graph m m m m D ...