8-2 Measuring Range and Accuracy
Temperature: 23±5˚C, humidity: 75% RH max., built-in battery
voltage 2.4 V or above.
rdg (reading): Read value, dgt (digit): Number of counts of last
digit
ACA
400.0A
1000A
Range
Accurac
Remarks
±
(
1.7%rdg+5dgt
)
±(1.7%rdg+5dgt
)
・
This meter is of average value response.
Frequency range: 50/60 Hz (Sinusoidal wave AC)
・
Accuracy is specified when an object is measured
at the center of the clamp sensor (CT),
・
A guide for a range of conductor positions
for accuracy guarantee
In other places of conductors to be
measured, the accuracy is as follows:
A area: 4% is added to the specified
accuracy.
B area: 1% is added to the specified
accuracy.
・
For accuracy when the indicated value is 15% of
the range or less, 8 dgt is added to the accuracy.
・
An induction error below 0.06 A/A may occur due to
influence of current flowing in adjacent conductors.
・
It may malfunction when measuring voltage/current in
the inverter circuit.
Center
A
B
400.0mV
4.000V
40.00V
400.0V
600V
Range
Accuracy
Input Impedance
Remarks
±(1.2%rdg+3dgt
)
±(1.9%rdg+3dgt
)
±(2.2%rdg+4dgt
)
Approx. 1000M
Ω
Approx. 10M
Ω
DCV