The cartridge contains many bubbles
The dosage selector cannot be turned
The pen has been stored with a needle attached
Solution:
- Always store your JuniorSTAR
®
pen without a needle attached�
- Perform the safety test until the large air bubbles are removed from
the cartridge� If large air bubbles remain in the cartridge, replace
the cartridge with a new one
The cartridge could be broken
Solution:
Replace the cartridge with a new one
The insulin cartridge does not contain enough insulin for your full
dose
Solution:
- EITHER: inject what is remaining and then complete your dose with
a new cartridge