- 6 -
6‑2 Main body (Rear view)
⑦Receptacle of the probe connector (PROBE)
・
The receptacle of the main body for connecting the cable connector
⑮
with the probe cable.
⑧
Output terminal (OUTPUT, RESET, GND)
・
The terminal is for connecting the no-voltage contact output, reset, GND to outside.
・
○
+
symbol on the both ends of the terminal are screws for fitting the terminal.
・
No 1,2 and 3 on the left side of the terminal are for the no-voltage contact output as
shown in right sketch.
No.1 is NC (Normally Close). Normally ON condition to No.2 COM
terminal excepting relay acting condition.
No.2 is COM (Common).
The common terminal to No.1 and No.3.
No.3 is NO (Normally Open). Normally OFF condition to No.2 COM
terminal excepting relay acting condition.
・
As the contact capacity of the relay is Max. AC 250V / DC 125V,
5A, use the Detector under these values.
・
When an outside load is inductive one, prevent the generation of the noises
in the varistor, diode and Sunaba circuit etc.
・
No.5 and No.6 are the terminal for operating the reset at a distance from the main body.
When use these terminals, remove the jumper fitting set in No.5 and No.6, and connect a 2-core
cord to the terminals and fit the switch etc. at the point of the cord.
The operating voltage of the circuit is 5V, use the switch having current capacity more than 1A.
The switch is a momentary type which returns original position by releasing the finger. Use the
COM and NC terminals which are normally ON position and become to OFF position by pressing.
When the jumper fitting is removed without using these terminals, the hold function can not be
operated even though the hold switch
④
ON.
When the terminals are not used, be sure to set the jumper fittings.
・
No.8 is the terminal for connecting the earth.