VIII. Simple Troubleshooting

Failure

Failure Cause

Approach

Machine turned on

but unelectrified

Power cord is not plugged

in place.

Check and ensure the power interface and the
power plug isplugged in place and in good
contact.

Fuse blown

Pull out the fuse box from the power supply
interface of the chiller, check the fuse, replace
with spare fuse if necessary, and check whether
the power supply voltage is stable. Check and
ensure the power interface and the power plug is
plugged in place and in good contact.

Flow Alarm, connecting

the wateroutlet and inlet

with a water pipe directly,

but still no water flow

Water level in the storage

water tank is too low

Check the water level gauge display, add water
until the level shown in the green area; And
check whether the water circulation pipe leaks.

When connecting the

chiller with the machine

to be cooled, flow alarm is

triggered; but when

connecting the water

outlet and inlet with a

water pipe directly, water

flows and the flow alarm

is not triggered.

Water circulation pipes
are blocked or a pipe is
bent or deformed

Check the water circulation pipe.

Ultra-high temperature

alarm

Blocked dust gauze, bad

thermolysis

Disassemble and clean the dust gauze regularly.

Poor ventilation for air

outlet and inlet

Ensure a smooth ventilation for air outlet and
inlet.

Voltage is extremely low

or unstable

Improve the power supply circuit or use a voltage
regulator.

Improper parameter

settings on thermostat

Reset the controlling parameters or restore the
factory setting.

Switch the power on and

off frequently

Ensure the time for the refrigeration process after
starting the chiller (at least 5 minutes).

Excessive heat load

Reduce the heat load or use other model with
larger cooling capacity.

Alarm for ultra-high room

temperature

The working ambient

temperature is too high for

the chiller

Improve ventilation and ensure that the machine
is operated below 40°C.

Serious problem of

condensate water

Water temperature is

much lower than ambient

temperature and the

ambient humidity is high

Increase water temperature or insulate pipes.

Water drains slowly from

outfall during water

changing

Water supply inlet is not

open

Open the water supply inlet

Содержание CW-5000 Series

Страница 1: ...TY JS 220308 004 1 CW 5000 5200 Industrial Circulating Chiller Installation Operation Instructions...

Страница 2: ...rly so that you can refer to it whenever you need information This manual is not a quality guarantee Our company reserves the right to the interpretation of the correction of misprint and improperly d...

Страница 3: ...TENTS Notice 4 I Overview 7 II Model Illustration 7 III Product Outlines Parts 8 IV Installation 10 V Operating States Parameters Adjustment 12 VI Alarm Output Terminal 14 VII Maintenance 15 VIII Simp...

Страница 4: ...It is not permitted to install the chiller in dangerous places with inflammable gas Electrical connection a The power supply must conform to the standard indicated in the nameplate or the manual b Use...

Страница 5: ...uideline Transportation with liquid is prohibited The equipment is not allowed to be transported with liquid to prevent internal pipeline leakage Transport a The equipment should be fixed firmly befor...

Страница 6: ...rease and the power consumption will increase and the alarm will not work properly Equipment surface cleaning a Please use cleaning agents that are non corrosive to metals and plastics b Please keep t...

Страница 7: ...chiller the refrigerant in the evaporator coil absorbs the heat of the return water and vaporizes into steam The compressor continuously extracts the generated steam from the evaporator and compresses...

Страница 8: ...TY JS 220308 004 8 III Product Outlines Parts TEMPERATUR E CONTROLLER POWER SWITCH AIR INLET FILTER GAUZE...

Страница 9: ...TY JS 220308 004 9 Note The above mentioned product outlines and part names are only for one of the models and they will be slightly different in different models...

Страница 10: ...Requirements Cooling medium allowed by the chiller includes purified water distilled water high purity water and other softened water It is forbidden to use oily liquids liquids containing solid part...

Страница 11: ...ical connection 1 Please select the cable based on the max rated current indicated on the label of the chiller 2 Recommended reference standard for power cord diameter selection Rated current A 5 10 1...

Страница 12: ...eeting equipment cooling requirements The new T 504D intelligent temperature controller is designed with constant temperature control mode with water temperature maintained at 25 Users can adjust it a...

Страница 13: ...alarm display remains until the alarm condition is eliminated 5 General settings adjustment Press SET button SET to enter the user defined state Meanwhile D1 flashes and indicates that the controller...

Страница 14: ...to 26 5 if the air temperature is up to 30 5 water temperature will be between 28 0 to 27 0 Case 2 The cooling water temperature is controlled by the constant mode It is required that the water temper...

Страница 15: ...larm output port is connected with a set of normally open or normally closed contactsof the relay inside the machine The working current is required to be less than 3A while the working voltage is les...

Страница 16: ...is not triggered Water circulation pipes are blocked or a pipe is bent or deformed Check the water circulation pipe Ultra high temperature alarm Blocked dust gauze bad thermolysis Disassemble and clea...

Страница 17: ...Shengshi Industrial Zone Zhongcun Town Panyu District Guangzhou Sales Company Shanghai Top Nine Industrial Co Ltd Seller s Company Address No 208 Wulian Road Room164 Shanghai Pudong New District China...

Отзывы: