64
2
3
4
1
5
7-5. Standby operation
[Fig. 7-10] Standby operation setting screen
It sets Y/N of standby operation.
It sets the temperature range to be applied for standby operation.
It sets the humidity range to be applied for standby operation.
It sets the standby time to be applied when the indicated data for
temperature or humidity does not enter into the standby operation
range (TEMP PV or HUMI PV).
●
It standbys indefinitely for entry to the standby operation range when
the standby operation time is set in “00.00.”
It decides either of “Entire” and “Maintain SEG” for standby operation method.
●
Entire : The standby operation is applied to the set entire segment in
[7-1 Program pattern setting]
●
Maintain SEG : The standby operation is applied only to the set
maintain range segment in [7-1 Program pattern setting]
①
②
③
⑤
④
● It is a screen to set the temperature&humidity Stand-by Range and Stand-by Time during Program operation.
● The set standby operation here is applied to [Fig. 7-1 Program pattern setting].
● Definition of standby operation
- Coditions for Stand-by Operation Entry (“OR” condition) : In case of no entry of either of temperature or humidity into the set range for standby operation
within the set segment time
- Conditions for Stand-by Operation Release (“AND” condition) : In case of entry of temperature and humidity into the set range for standby operation
- The standby time has indefinite value when the standby time is not set (Initial value).
Содержание TEMI1000 Series
Страница 1: ......
Страница 10: ...Part 02 Main screen 9...
Страница 37: ...Part 05 Appointed operation setting 36...
Страница 69: ...Part 08 Screen display setting 8 1 Setting display setting 69 8 2 DI error creation history view 71...
Страница 75: ...Part 09 Communication error 74...
Страница 78: ...MEMO...
Страница 79: ......
Страница 80: ......