12. Device Power Down and Removal
12.1. POWER OFF THE POWER CONTROLLER
Move the
Lock Switch
up to the ‘unlocked position’ (if locked) then press and hold the
Power Button
on the
sam
Power Controller
for at least 1 second
to turn the device
‘OFF’ (See Figure 11B for device diagram).
12.2. REMOVE DEVICE FROM SKIN
A.
Remove the bandage(s) from the skin.
B.
Remove the applicator from the
sam
Ultrasound Coupling Bandage by pulling
the tab on the side of the gel cup. This unlocks the cup from the applicator
(See Figure 12).
C.
Throw away the
sam
Bandages.
Warning: NEVER use a
sam
Ultrasound Coupling Bandage for more than one use.
Misuse of
sam
Ultrasound Coupling Bandages or use for more than the
intended treatment duration MAY RESULT IN BURN OR REPEATED SHUTOFF
of the applicator.
12.3. CLEAN THE APPLICATORS
Clean any residual ultrasound coupling media off of the applicator and off of the skin.
See section 14 for ‘Cleaning and Maintenance’ details.
12.4. RECHARGE THE DEVICE
Connect the
sam
Power Controller to the
sam
Electrical Charger for recharging. Prior
to the next use, allow up to 6 hours to recharge the battery to 100% charge. See section
9 for charging instructions.
MODEL
AT
-12
Pull Tab
fi gure 12: Removing Applicator