background image

Model  sam-12

Ultrasonic 

Diathermy 

Device

Directions 

for Use

Содержание sam-12

Страница 1: ...Model sam 12 Ultrasonic Diathermy Device Directions for Use ...

Страница 2: ... Setup Instructions 13 10 Treatment Options 13 11 Application Instructions 16 12 Device Power Down and Removal 22 13 Modes of Operation 23 14 Cleaning and Maintenance 24 15 Storage and Operating Conditions 24 16 Disposal of Waste Products 25 17 Appendix 26 A Symbols 26 B Specifications 27 C Notice of Compliance Calibration and Operational Period 28 D Technical Information 29 E Electrical Immunity a...

Страница 3: ... Ultrasound Coupling Bandages The specifications put forth in this manual were in effect at the time of publication 1 1 GENERAL SAFETY Thoroughly read and understand the precautionary and operating instructions before attempting to operate the sam Ultrasonic Diathermy Device Know the limitations and hazards associated with using any ultrasound device Observe the precautionary and operational decals ...

Страница 4: ...is might result 3 2 WARNINGS If the treatment is reported as painful or too hot at any point during treatment turn off device and remove the device from the skin Instruct the patient to inform the practitioner if the patient feels any pain or burning during treatment Instruct the patient how to turn off the sam Device and remove the sam Applicator if the patient feels any pain or burning during trea...

Страница 5: ...ls DO NOT apply directly over a bone that is near the skin surface 3 3 PRECAUTIONS Precaution should be taken when using the device Over an area of the spinal cord following a laminectomy i e when major covering tissues have been removed On patients with hemorrhagic diatheses Over areas where metal prosthesis or other metallic implants are embedded in tissue which may form a reflective surface to t...

Страница 6: ...ormance and safety of the device Do not use the device if any abnormal functionality occurs PRECAUTIONS Only recharge the sam Device using the sam Electrical Charger Use with any other recharging device may result in damage to the system and void all warranties When not in use power OFF the device to protect the functionality of the components Avoid dropping the applicator or power controller and ...

Страница 7: ...w 44 C during normal operation See section 13 4 on page 24 for more information 4 4 sam ULTRASOUND COUPLING BANDAGES The sam Device utilizes ultrasound coupling bandages which are manufactured with ultrasound coupling media sealed inside The ultrasound coupling bandages ARE REQUIRED to secure the sam Applicators to the body 4 5 BATTERY OPERATION Powered by a rechargeable lithium ion battery the sa...

Страница 8: ... 1 sam Model 12 Components AT 12 YB 12 if included AT 12 5 sam Components sam MODEL 12 COMPONENTS AT 12 Ultrasound Applicators UB 12 Ultrasound Coupling Bandages OM 12 User Manual PMC 12 Power Controller Module CT 12 Electrical Charger YB 12 Y branch Adapter PMC 12 UB 12 AT 12 YB 12 CT 12 AT 12 ...

Страница 9: ... Connection Bottom Lock Switch Indicator LED HRS DONE 20 1 0 0 80 60 40 3 2 1 3 5 4 2 5 1 5 figure 2 sam Interface Components 7 Accessories The sam Device may be used with any of the following accessories sam Belt Clip sam Armband sam Carrying Case Additional sam Ultrasound Coupling Bandages may be obtained by contacting the manufacturer 6 Operator Interface ...

Страница 10: ...ndicator LED Meaning Opaque absence of light Power Controller is OFF No power is being generated Blue Power Controller is ON Note Ultrasonic energy is generated only when the Power Controller is connected to the sam Applicator s Green Power Controller battery is fully charged Amber Power Controller battery is charging Red Too many applicators are attached A maximum of two applicators may be attach...

Страница 11: ...D Color Definitions Treatment Time Indicator LED Meaning Blue Treatment time remaining Treatment timer is set in one hour increments and will count down in 30 minute increments Amber Treatment complete This single done light will activate only when a treatment cycle has completed and will remain illuminated for 8 hours then turn off Table 3 Treatment Time Indicator LED Color Definitions 8 3 APPLICATO...

Страница 12: ...URATION The sam Ultrasonic Diathermy Device provides ultrasound therapy at a preset frequency and intensity The user can program the treatment duration to be 1 2 3 or 4 hours The maximum treatment duration setting on the sam Device is 4 hours Treatment duration should be set to the minimum increment required for effective therapy Due to individual differences in skin type and tolerance it is recomme...

Страница 13: ...over a bone that is near the skin surface 10 3 ONE OR TWO APPLICATORS The sam Device may be used with one or two applicators simultaneously The decision to use one or two applicators is dependent on the size and anatomical area of treatment Using two applicators allows a larger area of tissue to receive ultrasound treatment For example large anatomical regions such as the shoulder could benefit fro...

Страница 14: ...ors and two sam Ultrasound Coupling Bandages simultaneously the sam Ultrasound Coupling Bandages should be positioned so that their footprints do not overlap Caution DO NOT overlap sam Ultrasound Coupling Bandages Figure 6 Caution DO NOT use the Y adapter when using only one applicator for treatment Caution Do NOT attach more than one Y adapter to the sam Device Note If two applicators will be use...

Страница 15: ... lights Figure 7B The lights will illuminate to show how much battery life is remaining in the device If the device is not fully charged it may not be able to deliver treatment for 4 hours The device will only permit the user to set a treatment duration for which the battery charge is capable of fulfilling i To check the treatment duration for which the battery charge is capable of fulfilling simply...

Страница 16: ... other end of the Y adapter Figure 8 A Insert the power controller wire into the matching cavity at the base of the applicator at a 45 degree angle Figure 8A B Gently twist the wire jack clockwise until the applicator edge is flush with the wire jack Figure 8B HRS DON E 20 100 80 60 40 3 2 1 3 5 4 2 5 1 5 figure 8 Connecting Y Adapter Rotate Clockwise figure 8B Gently twist to Lock 45 figure 8A Insert...

Страница 17: ...el cup until a clicking noise is heard or clicking sensation is felt Figure 9C If using dual applicators repeat these steps with the second applicator and the second bandage Note When the applicator is pressed down onto the gel cup do not be concerned if coupling media flows out the edges of the gel cup Wipe any excess coupling media away with a tissue or towel Note Ensure the applicator is fully a...

Страница 18: ... the coupling media is completely uncovered ii Add more sam Coupling Media as needed B Turn the bandage over and adhere the bandage to the desired treatment location Figure 10B i The sam Coupling Media must be in direct contact with the skin If using dual applicators repeat these steps to adhere the second bandage and applicator to the second treatment site Note When applying bandages both sides o...

Страница 19: ...ON for about 4 seconds after turning ON the device iii The treatment timer LEDs and blue indicator LEDs power controller and applicator will remain ON for the duration of treatment C Press the Toggle Button to select 1 2 3 or 4 hour treatment duration as allowed by current battery charge Figure 11C i If the Toggle Button is pressed up to the 4 hour setting and a decreased treatment time is desired...

Страница 20: ...uration set by the user The treatment timer LEDs will count down in 30 minute increments as treatment progresses Upon completion the device will automatically shut off and the treatment complete LED will illuminate for approximately 8 hours before turning off See section 12 for how to manually power OFF the device Warning If the sam Lock Switch is in the locked position the sam Power Button will not...

Страница 21: ...ator See Figure 12 C Throw away the sam Bandages Warning NEVER use a sam Ultrasound Coupling Bandage for more than one use Misuse of sam Ultrasound Coupling Bandages or use for more than the intended treatment duration MAY RESULT IN BURN OR REPEATED SHUTOFF of the applicator 12 3 CLEAN THE APPLICATORS Clean any residual ultrasound coupling media off of the applicator and off of the skin See section ...

Страница 22: ...ment duration concludes 13 2 ON LOCKED MODE Once the device is powered ON the user may lock the control panel settings by sliding the Lock Switch on the sam Power Controller Module to the locked position as indicated by the lock symbol on the sam Power Controller Module This will prevent the device from unintentional shutoff during treatment While locked the Power Button and the Toggle Button canno...

Страница 23: ...other displays battery indicator treatment timer indicator LED and applicator LEDs will be inactive All ultrasound emissions from the applicator will cease The Treatment Complete LED will remain illuminated for approximately 8 hours after completion of therapy at which point it will time out and turn off To set the sam Device for a new treatment remember to move the lock switch to the unlocked posi...

Страница 24: ...rature 1 44 C Humidity 10 80 Atmospheric pressure range 700 1060 hPa Caution Do not keep the device in extreme hot or cold temperatures above 50 C or below 0 C Do not leave the sam Device in a hot or freezing car Do not leave the device in direct sunlight for extended periods UV light may damage or discolor the device 16 Disposal of Waste Products sam Ultrasound Coupling Bandages are one time use ...

Страница 25: ...adiation Separate collection for electrical and electronic equipment Must not be disposed of in unsorted municipal waste Caution consult accompanying documents Diverging beam Continuous wave CW Watts ultrasonic power Frequency in megahertz Effective radiating area Beam non uniformity ratio Use by date Temperature limits Humidity limitation Serial number Catalogue Reorder Number Lot number A SYMBOLS...

Страница 26: ...ividual Transducer Dimension 5 cm2 emitting surface area circular BNR 5 1 ERA 6 cm2 20 Maximum Treatment Duration 4 hours Other Electrical Ratings for the sam Device Components Electrical Charger Model GS2U 006 050 A Input Voltage 100 240 V 50 60 Hz Input Current 0 2 A Output Voltage 5 V DC Output Current 1 2 A Max Output Power 6 W Power Controller Output Voltage 3 7 V DC 10 Max Output Amperes 700...

Страница 27: ...ONIC INFRASONIC AND ULTRASONIC RADIATION EMITTING PRODUCTS The sam Device meets IEC 60601 1 3 ed 2012 IEC 60601 1 2 3 ed 2007 Medical Electrical Equipment General Requirements for Safety Electromagnetic Compatibility The sam Device is calibrated to provide 3 MHz 20 ultrasound diathermy at 0 65 W 20 per ultrasound applicator Operational Period The sam Power Controller Module is intended to maintain...

Страница 28: ...ct Support 203 349 2798 D TECHNICAL INFORMATION figure A Ultrasound Field Scan Across Applicator Face figure B Ultrasound Field Scan Away From the Applicator Face figure B Ultrasound Field Scan Away From the Applicator Face ...

Страница 29: ...ches 6 x 2 Guidance and Manufacturer s Declaration Emissions Emissions Test Compliance Electromagnetic Environment Guidance RF Emissions Group 1 The sam uses RF energy only for its internal CISPR 11 function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF Emissions Class B The sam is suitable for use in all establishments CISPR...

Страница 30: ...of a typical commercial or hospital environment Surge 1kV Differential 1kV Differential Mains power quality should be IEC 61000 4 5 2kV Common Common N A that of a typical commercial or no ground hospital environment Voltage 95 Dip for 95 Dip for Mains power quality should be Dips Dropout 0 5 Cycle 0 5 Cycle that of a typical commercial or IEC 61000 4 11 60 Dip for 60 Dip hospital environment If the...

Страница 31: ...rated from the sam by no less than the distances calculated listed below Conducted RF 3 Vrms V1 3 Vrms D 3 5 V1 Sqrt P IEC 61000 4 6 150 kHz to 150 kHz to 80 MHz 80 MHz Radiated RF 3V m D 3 5 Eq Sqrt P IEC 61000 4 3 80 MHz to E1 3 V m 80 to 800 MHz 2 5 GHz D 7 E1 Sqet P 800 MHz to 2 5 GHz Where P is the max power in watts and D is the recommended separation distance in meters Field strength from fi...

Страница 32: ... 80MHz 80 to 800MHz 800MHz to 2 5GHz D 3 5 V1 Sqrt P D 3 5 E1 Sqrt P D 3 5 E1 Sqrt P 0 01 0 11667 0 11667 0 23333 0 1 0 36894 0 36894 0 73785 1 0 1 1667 1 1667 2 3333 10 0 3 6894 3 6894 7 3785 100 0 11 667 11 667 23 333 F WARRANTY ZetrOZ Inc offers a 1 year manufacturer s warranty for the sam Device If the sam Device fails due to defects in material or workmanship ZetrOZ Inc at its discretion will ...

Страница 33: ...the power the Applicator does not controller wire jack and or the Y adapter wire jack If illuminate when the recently in sam Sensing Mode ensure that the applicator device is turned ON has had time to reach normal operating temperature and the indicator LED on the applicator has turned from red back to blue If the applicator LED still fails to illuminate blue contact the manufacturer with serial a...

Страница 34: ...losed loop continuous temperature monitoring which maintains treatment site temperatures below 44 C during normal operation In the event that the treatment site underneath the applicator reaches the temperature threshold the applicator will pause ultrasound output and vibrate once with a red LED notification that the device has begun a cooling or rest cycle sam will automatically resume treatment w...

Страница 35: ...sam Manufactured by ZetrOZ Inc 56 Quarry Road Trumbull CT 06611 USA Toll Free Tel 1 888 202 9831 USA User Manual OM 12 LL 2535 01 Rev B Published 06 04 2014 0086 ...

Отзывы: