For example: 

Hardness HRC converted from hardness L using impact device D should be written as 35, 

9 HRCLD. 

Where: 

35=Hardness value HL 

 

9=Hardness value HRC 

 

L=Leeb’s Method 

 

D=Impact device 

2.

 

Features and Applications 

2.1.

 

Introduction 

This  instrument  is  an  advanced  state-of-the-art  palm  sized  metal  hardness  tester  with 

many  new  features  which  are  light  weight,  easy  operation,  integrated  design,  high 
contrast display, low operating temperature, auto compensating for impact direction and 
etc. It can be widely used for measuring hardness of almost all ferrous and non-ferrous 

metal materials for scale of Leeb hardness, Rockwell C, B &A, Brinell, Vickers, Shore and 

Strength. 

It  has  a  memory  capacity  of  400  blocks  and  360000  data  which  can  be  downloaded  to 
computer via USB or Bluetooth (Optional). The measuring value can be printed out from 

tester to micro-printer by USB or Bluetooth. All stored data can be recalled and read on 

the tester easily.   

It  also  has  a  very  unique  feature,  which  impact  device  can  convert  between  D  and  DL 
simply  by  changing  impact  body.  This  two-in-one  probe  is  equivalent  to  two  individual 

probes. With this optional accessory, you can take measurement at very narrow surface 

such as slot bottom, gear tooth that probe D cannot match. 

The 3.7V Li-ion rechargeable battery inside the tester can be charged via USB from PC or 
via  individual  battery  charger  from  mains  wall  power.  With  data  software  for  PC, 

customers can  download  measuring  values  from  THE  TESTER  to  PC and  make  process 

such as save, delete, create testing report and export them to Excel. 

2.2.

 

Specifications 

 

Principle: Leeb hardness measurement 

 

Accuracy: +/-2HL    (or 0.3%@HL=800) 

 

Display: Digital with high contrast OLED 

 

Display mode: Normal/flip or upward/downward 

 

Hardness scale: HL / HRC / HRB / HB / HV / HS / HRA / σb 

 

Measuring  range:  HL170-960  /  HRC19-70  /  HRB13-109  /  HB20-665  /  HV80-940  / 

HS32-99.5 / HRA30-88 

 

Impact device: D / D-DL 2 in 1    (Optional) 

 

Materials: 11 common metal materials 

 

Memory: 400 blocks and 360000 data can be saved and re-readable   

 

Statistics: Average/Max/Min value can be calculated automatically 

 

Interface:  USB/RS232  for  data  transferring,  printing  or  charging  battery,  Bluetooth  for 

data transferring and printing 

 

Recalibration: allowed by user 

 

Alarm: Up or down limit 

 

Indicator: Low battery 

 

Power supply: 3.7V Li-ion rechargeable battery 

 

Power on/off: Auto 

 

Operating environment: -40⁰C ~+70⁰C 

 

Dimension (LxWxD): 148mm×44mm×22mm 

 

Net weight: 110g 

 

Standard: ASTM A956 

Содержание HardyTest HARTIP 1800

Страница 1: ...HARDNESS TESTER HardyTest HARTIP 1800 OPERATION MANUAL...

Страница 2: ......

Страница 3: ...SAMPLE OF CURVING SURFACE 7 5 6 SUPPORTING THE SAMPLES DURING TESTING 7 5 7 SAMPLES WITH CURVED SURFACES 7 6 OPERATION 8 6 1 BUTTON DESCRIPTION 8 6 2 DIAGRAM OF OPERATION 9 6 3 POWER ON THE INSTRUMEN...

Страница 4: ...ional to the velocities of the magnet Leeb hardness is expressed by the following formula 1000 Vi Vr HL Where HL is Leeb Hardness Vr is the rebound velocity of the impact body Vi is the impact velocit...

Страница 5: ...l probes With this optional accessory you can take measurement at very narrow surface such as slot bottom gear tooth that probe D cannot match The 3 7V Li ion rechargeable battery inside the tester ca...

Страница 6: ...Recalibration allowed by user 2 4 Applications Hardness tests on installed machines or steel structures e g on heavy and large work piece or on permanently installed system parts Rapid testing of mult...

Страница 7: ...MATERIALS H R C H R B H B H V H S HRA b N mm2 STEEL CAST STEEL 20 0 67 9 59 6 99 5 80 647 80 940 32 5 99 5 30 88 375 1710 ALLOY TOOL STEEL 20 5 67 1 80 898 1170 2639 STAINLESS STEEL 19 6 62 4 46 5 101...

Страница 8: ...direction of impact should be perpendicular to the coupling surface For the coupling operation the following prerequisites must be fulfilled The contact surface of the sample and the surface of the b...

Страница 9: ...be checked in the following manner a Medium weight samples and also heavier samples with protruding parts or thin walls should be placed on a solid support in such a manner that they do not move or fl...

Страница 10: ...start to print a new page while Bluetooth printing is on Delete current reading after set up statistics Upward button Move cursor upward Change value or digit circulative from 0 to 9 Shift normal or u...

Страница 11: ...9 6 2 Diagram of Operation Figure 6 2 Diagram of Operation...

Страница 12: ...s required press the confirmation button on the keyboard to enter main menu Figure 6 4 Main Menu After entering main menu press or button to move among the different options press button to enter conf...

Страница 13: ...t up Impact device to DL while you are using probe DL Figure 6 6 Impact Device Selection After selecting impact device Successfully will be displayed on the screen then return to the Parameters menu F...

Страница 14: ...Different hardness scales will show on the screen depending on different materials The default hardness scale is Leeb hardness value HLD Figure 6 10 Hardness Scale 6 4 5 Impact Direction Choose Impact...

Страница 15: ...tatistics menu and press button to enter sub menu Then press or button to choose different mean times then press button to confirm the selection and return to statistics Menu The default mean time is...

Страница 16: ...s button to confirm the selection and return to statistics menu The default value of upper limit alarm is 999 HLD When the users set the value of upper limit alarm the ranges of hardness values corres...

Страница 17: ...t as any block of existing blocks Press button to move the cursor and choose the value of modified group No cycling from hundreds tens units to return in turn When the cursor is moved to the option of...

Страница 18: ...window All blocks stored in the instrument are showed on the interface At most 9 block numbers are showed on each screen Change Page is showed on upper left corner Displayed content can be changed by...

Страница 19: ...irm Figure 6 24 Communicating by USB RS232 Figure 6 25 Communicating by Bluetooth Via USB RS232 or Bluetooth interface testing values can be downloaded to PC or printed by micro printer with our speci...

Страница 20: ...nsferring on line Start software on PC and operate the relative tool to download the data from memory Note Please make sure the connection between PC and tester is successful See 11 3 6 7 4 Print bloc...

Страница 21: ...s hardness scale Under Bluetooth communication mode pressing button In order to meet the measurement of requiring different direction it is also designed to change display direction which function can...

Страница 22: ...apacity and storage capacity are showed on the screen After viewing them press button to return to the function menu Figure 6 32 System info 6 8 8 User info Choose User info from the function menu and...

Страница 23: ...vice D and DL should be operated separately Before calibration you should choose impact device from parameters menu first and then perform calibration for this impact device 6 8 10 Brightness Adjustme...

Страница 24: ...e measurement at very narrow surface such as slot bottom gear tooth that probe D cannot match Please refer to the following steps to change impact body Unscrew the support ring from the tester and let...

Страница 25: ...oper way of holding is important for obtaining better readings Attention the tester must be placed against object surface firmly and perpendicularly A slight gap between support ring of tester and sur...

Страница 26: ...some time Please make sure to obtain additional or replacing battery from SALUTRON Otherwise it may cause the instrument to get un accuracy value In order to avoid damage of battery due to over discha...

Страница 27: ...mode 9 3 System reset The instrument doesn t response possibly due to instable power voltage or others the instrument can be reset when it does not response Take the impact body out of the instrument...

Страница 28: ...ndrical HZ 11 13 20 18 5mm R 11mm 13mm R 11mm not possible R 30mm D6a HZ 12 5 17 20 20 5mm R 12 5mm 17mm HZ 16 5 30 20 20 5mm R 16 5mm 30mm spherical K 10 15 20 7 7mm R 10mm 13mm R 10mm not possible R...

Страница 29: ...wer indicator green light keep bright when the battery power 50 30 three color power indicator the yellow light keep bright when the battery is less than 30 three color power indicator the red light k...

Страница 30: ...s the Drive Letter of the application CD x cp2102 Drivers CP210x USB Composite Device x cp2102 Drivers CP210x USB to UART Bridge Controller Install these two drivers 11 2 Installation of Software To r...

Страница 31: ...nstalled during installation install the file NET35_SP1_cn exe that is located in X dotnet_35 folder in the CD You can also download it from the Microsoft website 11 3 Operation of PC Software Double...

Страница 32: ...which is in the down left corner of the software window After checking an orange check displays that means the connection is created successfully Before Checking Device After Checking Device connecte...

Страница 33: ...saved 11 3 4 Test data Click Test data tab in the working area or click button from the tool bar to load the Test data working area After connection is established you can return to the testing mode...

Страница 34: ...isplay choose the file and click Open the saved data will be displayed in the current data area File Print to print current data Select Print from the File menu or click button the data in current win...

Страница 35: ...click Read button to read the current parameters from the tester select new parameters in Preinstall area and click Modify button to change the current parameters Note Operations in Parameter tab must...

Страница 36: ...l value of standard block in the standard value box and press Change standard value The default value is 500 The modified data is saved in the application 2 Calibration To take measurement on the stan...

Страница 37: ...ew form Click button to open saved stored data Click button to read stored data Click button to save current data into Microsoft Excel document Click button to print current data to the default printe...

Страница 38: ...36 11 4 Quit the program After finish all the job you can click from the tool bar to quit the program...

Страница 39: ......

Страница 40: ...tf S ri a e l d u C T r r e o m n er M A e ll s e s e t 3 e 0 c h 7 n ik D 5 G 0 m 22 b 6 H Dr Got Frechen Tel 49 0 2234 9999960 Fax 49 0 2234 9999962 Email info salutron de www salutron de...

Отзывы: