![Saluki SE2031 Скачать руководство пользователя страница 85](http://html1.mh-extra.com/html/saluki/se2031/se2031_user-manual_3700286085.webp)
Tel: 886. 909 602 109 Email: [email protected]
www.salukitec.com
85
13. Display Harmonic Waveform
Choose the “Harmonic Wave” option in left window of the software.
Now there are two XY coordinates charts to display harmonic. Users can choose to set it to display R/X/Y/
θ
for each
harmonic, as is shown in Fig.70.
Fig.70 The Harmonic Waveform Display
5.3 Usage Examples
This example will show how to use SE2031 PC software to configure parameters, and observe and record the value
of R/X/Y/
θ
.
First, connect the SE2031 and PC according to the software usage guidance in 5.2 and then you can start to
configure.
Suppose that the lock-in amplifier configurations are shown in Table 14.
Table 14. Example Configuration Options
Input Source
Single-ended voltage input
Input Amplitude
40mV
Input Coupling
AC
Input Notch Filter
Close
Dynamic Reserve
Low Noise
Sensitivity
Auto Sensitivity
Reference Source
External reference, 1000Hz
Reference Signal Trigger Way
TTL rising edge trigger
Phase Shift Angle
0