Tel: 886.2.2175 2930 Email: [email protected]
31
When the FFT spectrum is being displayed, the channel vertical knobs become the zoom and position controls
corresponding to their respective channels. The VOLTS/DIV knob provides the following zoom factors: X1(default),
X2, X5 and X10. The FFT spectrum is magnified vertically to the marker M (math waveform reference point on the
left edge of the screen). Turn the VERTICAL POSITION knob clockwise to move up the spectrum.
5.3.1.7
Using Cursors to Measure FFT Spectrum
You may use cursors to take two measurements on the FFT spectrum: amplitude (in dB) and frequency (in Hz).
Amplitude is referenced to 0db that equals 1VRMS here. You may use cursors to measure at any zoom factor.
Push the CURSOR button, choose the Source option and then select Math. Press the Type option button to select
between Amplitude and Frequency. Click the SELECT CURSOR option to choose a cursor. Then use the V0 knobs
to move Cursor S and Cursor E. Use the horizontal cursor to measure the amplitude and the vertical cursor to
measure the frequency. Now the display at the DELTA menu is just the measured value, and the values at Cursor S
and Cursor E.
Delta is the absolute value of Cursor S minus Cursor E.
5.4
Trigger Controls
The trigger can be defined through the Trigger Menu and front-panel controls. There are six types of trigger: Edge,
Video, Pulse Width, Alter, Slope and Overtime. Refer to the following tables to find a different set of options for each
type of trigger.