Document Reference: ENG-UMAN-000731
Revision: 5.00
Page 28 of 61
Note: The arrow will change orientation to an Up arrow
for increasing stimulation
current, and various areas of the screen will become active, e.g. the ECAP Measurement
and Current rolling plot.
2.
Touch the Up arrow
to increase stimulation current to the desired level (mA)
3.
Touch the Down arrow
to decrease stimulation current (Figure 8.12G).
Note: The current will increase/decrease by the Current Increment (mA) set. The increment
in the current scale is shown on the right side (Figure 8.12D). The scale will extend to the
Maximum current set (Figure 8.12B).
A: Up arrow
B: Maximum current
C: Current scale
D: Current increment scale
E: Desired current level
F: Current increment
G: Down arrow
Figure 8.12: Features of the Stimulation Amplitude Control
8.8.1
Stop
1.
Touch the Stop button
to stop stimulation at any time.
8.9
Set up ECAP measurement
The stimulator records the ECAP signal and displays a graph of the ECAP on the
programming screen (Figure 8.13). ECAP measurement is only possible when the
stimulation is running and a clear signal is visible. Patients will usually be able to feel
paresthesia when a signal is detectable.
A
B
E
G
F
D
C