PRO

check 

®

 

Ultrasonic Oxygen Indicator

 

INSTRUCTIONS FOR CHECKING O

2

CONCENTRATORS

 

To turn unit on and off: 

To turn the unit on or off press and release the 

on/off

” 

button on the lower left-hand side of the unit. (EXAMPLE: 

Concentrator O

20.9) When the unit is turned on the device defaults to the 

mode in which the software enables the unit to measure the output of an 
oxygen concentrator compensating for the argon that passes through the 
sieve beds. The display will read 

"PRO"

, then 

"O

2

and finally 

"Concentrator O2" 

and a number.  If the 

PRO

check

® 

unit is hooked up 

to  an operating concentrator, the device will automatically measure the 
concentration of the oxygen being produced and show the value on the  
LCD display. The unit's default operating mode automatically compensates 
for the argon that passes through the sieve beds, it will display a false 
reading (

approximately 90%

) if connected to an oxygen source other than

 

a concentrator unless the specific actions discussed on 

pages 6 - 7 

of this 

manual are taken. If the device is not hooked up to an oxygen source it will 
measure and display the concentration of oxygen in the ambient air 
(

usually 20.9%

). 

Reading may vary if sample gas remains in unit

.

 

To turn the 

PRO

check

® 

device off, press and release the 

on/off

” 

button.  The unit will display 

“OFF” 

and shut down. To conserve battery 

life, it  automatically shuts down if any one of the unit's functions is not 
utilized  for approximately 5 minutes. 

Measuring Oxygen Concentrations:

 

Follow the steps below to measure the concentration of oxygen being 
supplied by an oxygen concentrator:

 

1.

 

Make sure the concentrator has been running long enough (follow 
concentrator manufacturer's guidelines) to reach full operational output.

 

2.

 

Make sure flow rate is above 2 LPM - preferably at the concentrator's 
highest output.

 

3.

 

Remove any humidifier from concentrator

.

 

4.

 

Connect supply tubing to the concentrator oxygen outlet and the inlet 
port on the 

PRO

check

®

.

 

5.

 

Turn on the 

PRO

check

® 

device (See above).

 

6.

 

The indicator will stabilize in about 10 seconds and provide an 
accurate reading on the LCD display.

 

7.

 

If there is any question to the accuracy of the reading, perform the 
"Pure Oxygen Verification Procedure" and if necessary the “Quick 
One Step Pure Oxygen Calibration Procedure on 

page 6

.

 

CAUTION: Do not check concentrator with humidifier in place. 

Humidity can affect unit's readings and damage the unit.

 

NOTES: Never check a concentrator while holding down the "pure O

/calibrate"

button. 

You will get a false reading. You must hold down the 

"pure O

/calibrate" 

button to 

check unit against a source of U.S.P. oxygen (> 99.0%) cylinder or liquid oxygen.

 

 

Содержание RO2 check

Страница 1: ...J CAUTION 30 Spur Dr El Paso TX 79906 Phone 1 800 421 0024 Email customercare salterlabs com www salterlabs com...

Страница 2: ...30 Spur Dr El Paso TX 79906 Phone 1 800 421 0024...

Страница 3: ...Liquid Oxygen Vessels or Oxygen Cylinders 5 VI Verification Calibration Instructions 6 Pure Oxygen Verification Procedure 6 Quick One Step Pure Oxygen Calibration 6 Full Two Step Calibration Procedure...

Страница 4: ...tal device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference in a residential installation This equipment generates uses and...

Страница 5: ...ff button will turn the unit on or off 2 The pure O2 calibrate button when pressed and held down while connected to a compressed gas or liquid U S P oxygen 99 0 source verifies that the unit is proper...

Страница 6: ...ncentration measurement mode can be quickly and easily verified by following the Pure Oxygen Verification procedure on page 6 of this manual 3 If accreditation organizations or internal policies and p...

Страница 7: ...2 check device off press and release the on off button The unit will display OFF and shut down To conserve battery life it automatically shuts down if any one of the unit s functions is not utilized f...

Страница 8: ...O2 check unit on 4 When unit stabilizes the unit display will read Concentrator O2 91 8 approx concentration 5 Press and continue to hold down the pure O calibrate button 6 Unit will display Pure O2 a...

Страница 9: ...l display Pure O2 and a numeric reading 7 If the unit is properly calibrated the display should read between 99 and 101 The PRO2 check is ready to use 8 If the unit display reads err U or any other re...

Страница 10: ...elease Calibration Pure O2 button 6 You will see three dashed lines scroll in circles Then the scrolling will stop and then you will see Cal End and read 91 7 Note Reading may range from 91 to 92 This...

Страница 11: ...ted on the rear of the unit away from the case Connect the 9 volt battery to the standard 9 volt battery connection and lay the battery sideways in the battery compartment Insert battery door by slidi...

Страница 12: ...ease verify calibration if used over 5 000 feet Note If readings do not seem accurate recalibrate the unit Vibration It is recommended that the unit be used in a stable position to prevent the display...

Страница 13: ...hargeable type battery is not recommended Operating Temperature 32 F to 105 F 0 C to 41 C Storage Temperature 29 F to 160 F 34 C to 71 C Specifications Concentration Mode Measured Oxygen Concentration...

Страница 14: ...ure that the unit is holding calibration against a U S P oxygen 99 0 source the unit does not require calibration My accreditation organization as well as our internal policies require periodic calibr...

Страница 15: ...pprox What s wrong Nothing When in the Concentrator Mode the electronic components of the indicator automatically compensates for the argon found in the gas produced by an oxygen concentrator As a res...

Страница 16: ...nd continue to hold down pure O2 calibrate button to temporarily remove argon correction from unit s With unit still connected to source of U S P oxygen Display reads between 98 to 102 but not 99 9 on...

Страница 17: ...n Manual Display shows err during use Unit has made a reading outside normal ranges Check calibration Make sure gas being measured is from oxygen concentrator LO BATT appears in display screen when un...

Страница 18: ...be responsible for all shipping charges to return the unit to Salter Labs Salter Labs will be responsible for normal shipping charges for the return of the unit to purchaser All warranties are voided...