background image

 

  

PRO

check 

®

 

Ultrasonic Oxygen Indicator

 

 

INTRODUCTION

 

Thank you for purchasing a 

PRO

check 

® 

Ultrasonic Oxygen Indicator

. The 

designers of this product have years of experience in the field of oxygen 
concentrator repairs.  The unreliability and high cost of operation of fuel 
cell analyzers led to the invention of the 

PRO

check 

® 

Ultrasonic Oxygen

 

Indicator

 

By relying on a method already used in many oxygen concentrators to 
trigger "Low O

2

" alarms – 

ultrasound 

– Salter Labs has produced a more 

cost effective product to check the output of oxygen concentrators. The 

PRO

check 

®  

Ultrasonic Oxygen Indicator 

responds faster, has no fuel 

cell  to replace and is simpler to use than a fuel cell analyzer. 

 

Directions for Use:

 

There are two (2) controls for the 

PRO

check

®  

unit: 

 

1.

 

Pressing and releasing the 

on/off

” 

button will turn the unit on or off. 

2.

 

The 

pure O

calibrate

” 

button when pressed and held down 

while connected to a compressed gas or liquid U.S.P. oxygen 
(> 99.0%) source verifies that the unit is properly calibrated, thus 
eliminating the need to have an outside source calibrate the unit. 

 

Since the 

PRO 

check

®  

device uses ultrasound technology to check the con- 

centrator operation, there has to be two modes of operation: 

 

1.

 

When the unit is turned on, the software enables the unit to 
measure the output of an oxygen concentrator adjusting for the 
argon that  passes through the sieve beds. 

 

 

2.

 

When using the unit with the 

pure O

calibrate

” 

button 

depressed the software measures oxygen not adjusting for the 
argon. When the 

O

calibrate

” 

button is released, the unit 

automatically returns to the concentrator mode and will then again 
adjust for the argon and read incorrectly for pure oxygen. 

 

We are sure that the 

PRO 

check

®  

Ultrasonic Oxygen Indicator 

will 

provide years of fast, reliable operation. 

 

 

 

Содержание RO2 check

Страница 1: ...J CAUTION 30 Spur Dr El Paso TX 79906 Phone 1 800 421 0024 Email customercare salterlabs com www salterlabs com...

Страница 2: ...30 Spur Dr El Paso TX 79906 Phone 1 800 421 0024...

Страница 3: ...Liquid Oxygen Vessels or Oxygen Cylinders 5 VI Verification Calibration Instructions 6 Pure Oxygen Verification Procedure 6 Quick One Step Pure Oxygen Calibration 6 Full Two Step Calibration Procedure...

Страница 4: ...tal device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference in a residential installation This equipment generates uses and...

Страница 5: ...ff button will turn the unit on or off 2 The pure O2 calibrate button when pressed and held down while connected to a compressed gas or liquid U S P oxygen 99 0 source verifies that the unit is proper...

Страница 6: ...ncentration measurement mode can be quickly and easily verified by following the Pure Oxygen Verification procedure on page 6 of this manual 3 If accreditation organizations or internal policies and p...

Страница 7: ...2 check device off press and release the on off button The unit will display OFF and shut down To conserve battery life it automatically shuts down if any one of the unit s functions is not utilized f...

Страница 8: ...O2 check unit on 4 When unit stabilizes the unit display will read Concentrator O2 91 8 approx concentration 5 Press and continue to hold down the pure O calibrate button 6 Unit will display Pure O2 a...

Страница 9: ...l display Pure O2 and a numeric reading 7 If the unit is properly calibrated the display should read between 99 and 101 The PRO2 check is ready to use 8 If the unit display reads err U or any other re...

Страница 10: ...elease Calibration Pure O2 button 6 You will see three dashed lines scroll in circles Then the scrolling will stop and then you will see Cal End and read 91 7 Note Reading may range from 91 to 92 This...

Страница 11: ...ted on the rear of the unit away from the case Connect the 9 volt battery to the standard 9 volt battery connection and lay the battery sideways in the battery compartment Insert battery door by slidi...

Страница 12: ...ease verify calibration if used over 5 000 feet Note If readings do not seem accurate recalibrate the unit Vibration It is recommended that the unit be used in a stable position to prevent the display...

Страница 13: ...hargeable type battery is not recommended Operating Temperature 32 F to 105 F 0 C to 41 C Storage Temperature 29 F to 160 F 34 C to 71 C Specifications Concentration Mode Measured Oxygen Concentration...

Страница 14: ...ure that the unit is holding calibration against a U S P oxygen 99 0 source the unit does not require calibration My accreditation organization as well as our internal policies require periodic calibr...

Страница 15: ...pprox What s wrong Nothing When in the Concentrator Mode the electronic components of the indicator automatically compensates for the argon found in the gas produced by an oxygen concentrator As a res...

Страница 16: ...nd continue to hold down pure O2 calibrate button to temporarily remove argon correction from unit s With unit still connected to source of U S P oxygen Display reads between 98 to 102 but not 99 9 on...

Страница 17: ...n Manual Display shows err during use Unit has made a reading outside normal ranges Check calibration Make sure gas being measured is from oxygen concentrator LO BATT appears in display screen when un...

Страница 18: ...be responsible for all shipping charges to return the unit to Salter Labs Salter Labs will be responsible for normal shipping charges for the return of the unit to purchaser All warranties are voided...