background image

 

PRO

check 

®    

 

Ultrasonic Oxygen Indicator

 

 

FULL "TWO STEP" CALIBRATION PROCEDURE

 

 

 

The following procedure may be used to calibrate the 

PRO

check

®  

indicator 

to both room air and U.S.P. oxygen (> 99.0%): 

 

1.

 

Clear analyzer gas chamber by 

inhaling 

through inlet. Use clean tubing and 

discard after use. 

2.

 

With analyzer off, Push and Hold 

Calibration/Pure O

button while turning on power. 

3.

 

When three dashed lines appear (---) release 

Calibration/Pure O

button. Display 

should read 

Cal Air 

(Cal Air). 

 

 

 

NOTE: Do not exhale breath into unit. Moisture will affect calibration; a slight suction 
can be used to evacuate the sampling chamber.

 

 

 

 

4.

 

Press and release 

Calibration/Pure O

button. You will see three dashed lines (---) 

scroll in circles. Then the scrolling will stop and you will see 

CAL O

2

 

5.

 

Connect to 

O

Cylinder at 2 LPM, then press and release 

Calibration/Pure O

button. 

6.

 

You will see three dashed lines (---) scroll in circles. Then the scrolling will stop and 

then you will see 

Cal End 

and read 91.7 

Note: Reading may range from 91 to 92. 

This is not a concern.

 

7.

 

Push the 

Calibration/Pure O

button. You should get a reading between 99% and 101%. 

8.

 

Unit is calibrated and ready for use. 

9.

 

If step 7 above is not achieved, try one more time. If 99% to 101% is still not achieved, 
call Salter Labs

® 

Customer Service to send unit back for service by a Technician at 

Salter Labs

® 

Service Dept. 

 

 

NOTE: If unit displays "CAL ERR" during verification or calibration 
procedures, check to ensure that you are using U.S.P. oxygen (> 99.0%) and 
that the flow is turned on to at least 2 LPM, then repeat procedure.

 

 

 

NOTE: If unit displays "---" during verification or calibration procedures, release 

the 

"pure O2/calibrate

” 

button or the unit will shut off automatically. 

 

 

If any problems are encountered during the calibration sequences, 
contact Salter Labs at 1-800-421-0024 or 1-661-854-3166. 

 

 

 

 

 

 

 

Содержание RO2 check

Страница 1: ...J CAUTION 30 Spur Dr El Paso TX 79906 Phone 1 800 421 0024 Email customercare salterlabs com www salterlabs com...

Страница 2: ...30 Spur Dr El Paso TX 79906 Phone 1 800 421 0024...

Страница 3: ...Liquid Oxygen Vessels or Oxygen Cylinders 5 VI Verification Calibration Instructions 6 Pure Oxygen Verification Procedure 6 Quick One Step Pure Oxygen Calibration 6 Full Two Step Calibration Procedure...

Страница 4: ...tal device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference in a residential installation This equipment generates uses and...

Страница 5: ...ff button will turn the unit on or off 2 The pure O2 calibrate button when pressed and held down while connected to a compressed gas or liquid U S P oxygen 99 0 source verifies that the unit is proper...

Страница 6: ...ncentration measurement mode can be quickly and easily verified by following the Pure Oxygen Verification procedure on page 6 of this manual 3 If accreditation organizations or internal policies and p...

Страница 7: ...2 check device off press and release the on off button The unit will display OFF and shut down To conserve battery life it automatically shuts down if any one of the unit s functions is not utilized f...

Страница 8: ...O2 check unit on 4 When unit stabilizes the unit display will read Concentrator O2 91 8 approx concentration 5 Press and continue to hold down the pure O calibrate button 6 Unit will display Pure O2 a...

Страница 9: ...l display Pure O2 and a numeric reading 7 If the unit is properly calibrated the display should read between 99 and 101 The PRO2 check is ready to use 8 If the unit display reads err U or any other re...

Страница 10: ...elease Calibration Pure O2 button 6 You will see three dashed lines scroll in circles Then the scrolling will stop and then you will see Cal End and read 91 7 Note Reading may range from 91 to 92 This...

Страница 11: ...ted on the rear of the unit away from the case Connect the 9 volt battery to the standard 9 volt battery connection and lay the battery sideways in the battery compartment Insert battery door by slidi...

Страница 12: ...ease verify calibration if used over 5 000 feet Note If readings do not seem accurate recalibrate the unit Vibration It is recommended that the unit be used in a stable position to prevent the display...

Страница 13: ...hargeable type battery is not recommended Operating Temperature 32 F to 105 F 0 C to 41 C Storage Temperature 29 F to 160 F 34 C to 71 C Specifications Concentration Mode Measured Oxygen Concentration...

Страница 14: ...ure that the unit is holding calibration against a U S P oxygen 99 0 source the unit does not require calibration My accreditation organization as well as our internal policies require periodic calibr...

Страница 15: ...pprox What s wrong Nothing When in the Concentrator Mode the electronic components of the indicator automatically compensates for the argon found in the gas produced by an oxygen concentrator As a res...

Страница 16: ...nd continue to hold down pure O2 calibrate button to temporarily remove argon correction from unit s With unit still connected to source of U S P oxygen Display reads between 98 to 102 but not 99 9 on...

Страница 17: ...n Manual Display shows err during use Unit has made a reading outside normal ranges Check calibration Make sure gas being measured is from oxygen concentrator LO BATT appears in display screen when un...

Страница 18: ...be responsible for all shipping charges to return the unit to Salter Labs Salter Labs will be responsible for normal shipping charges for the return of the unit to purchaser All warranties are voided...

Отзывы: