background image

 

PRO

check 

®    

 

Ultrasonic Oxygen Indicator

 

 

FREQUENTLY ASKED QUESTIONS

 

 

Thank you for purchasing the 

PRO

check

® 

Ultrasonic Oxygen Indicator

It  will give you years of trouble-free operation. The following are some of  
the frequently asked questions about the operation of this device. 

 

Do I have to calibrate my 

PRO

check

® 

unit daily?

 

No. The 

PRO 

check

® 

indicator was calibrated at the factory. 

 

How do I know that my 

PRO

check

® 

device is accurately 

measuring concentrations?

 

To verify, simply connect the unit to a source of either compressed gas or 
liquid U.S.P. oxygen (> 99.0%) at 2 LPM.  Then turn on the device and 
hold down the 

pure O

2

/calibrate

” 

button. As long as the display 

reads between 

99% and 101%

, the unit is properly calibrated. 

 

How often should I calibrate my 

PRO

check

® 

unit?

 

As long as you can verify 

(via the Pure Oxygen Verification Procedure) 

that the unit is holding calibration against a U.S.P. oxygen (> 99.0%) 
source, the unit does not require calibration. 

 

My accreditation organization as well as our internal policies require 
periodic calibration of this type of equipment. Can I calibrate the 

PRO

check

® 

device to comply with these requirements?

 

Yes, if accreditation organizations or internal policies and procedures 
require routine calibration, instructions on calibrating the device are 
found on 

pages 6 - 7 

of this manual.

 

 

Does the unit have to be turned off before verifying or calibrating? 

Yes. The unit must be turned off and evacuated to room air before you 
begin the oxygen verification or calibration procedures.

 

 

Why doesn't the unit drop immediately to room air?

 

Oxygen is heavier than room air, so it will linger inside the unit.  When 
there is a flow (either from room air or another concentrator), the reading 
will change immediately. 

 

Can I exhale (blow) into the unit to reach room air?

 

No.  You may apply a small suction or flow of room air, but do not exhale 
into the unit. 

 

 

 

11

 

Содержание RO2 check

Страница 1: ...J CAUTION 30 Spur Dr El Paso TX 79906 Phone 1 800 421 0024 Email customercare salterlabs com www salterlabs com...

Страница 2: ...30 Spur Dr El Paso TX 79906 Phone 1 800 421 0024...

Страница 3: ...Liquid Oxygen Vessels or Oxygen Cylinders 5 VI Verification Calibration Instructions 6 Pure Oxygen Verification Procedure 6 Quick One Step Pure Oxygen Calibration 6 Full Two Step Calibration Procedure...

Страница 4: ...tal device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference in a residential installation This equipment generates uses and...

Страница 5: ...ff button will turn the unit on or off 2 The pure O2 calibrate button when pressed and held down while connected to a compressed gas or liquid U S P oxygen 99 0 source verifies that the unit is proper...

Страница 6: ...ncentration measurement mode can be quickly and easily verified by following the Pure Oxygen Verification procedure on page 6 of this manual 3 If accreditation organizations or internal policies and p...

Страница 7: ...2 check device off press and release the on off button The unit will display OFF and shut down To conserve battery life it automatically shuts down if any one of the unit s functions is not utilized f...

Страница 8: ...O2 check unit on 4 When unit stabilizes the unit display will read Concentrator O2 91 8 approx concentration 5 Press and continue to hold down the pure O calibrate button 6 Unit will display Pure O2 a...

Страница 9: ...l display Pure O2 and a numeric reading 7 If the unit is properly calibrated the display should read between 99 and 101 The PRO2 check is ready to use 8 If the unit display reads err U or any other re...

Страница 10: ...elease Calibration Pure O2 button 6 You will see three dashed lines scroll in circles Then the scrolling will stop and then you will see Cal End and read 91 7 Note Reading may range from 91 to 92 This...

Страница 11: ...ted on the rear of the unit away from the case Connect the 9 volt battery to the standard 9 volt battery connection and lay the battery sideways in the battery compartment Insert battery door by slidi...

Страница 12: ...ease verify calibration if used over 5 000 feet Note If readings do not seem accurate recalibrate the unit Vibration It is recommended that the unit be used in a stable position to prevent the display...

Страница 13: ...hargeable type battery is not recommended Operating Temperature 32 F to 105 F 0 C to 41 C Storage Temperature 29 F to 160 F 34 C to 71 C Specifications Concentration Mode Measured Oxygen Concentration...

Страница 14: ...ure that the unit is holding calibration against a U S P oxygen 99 0 source the unit does not require calibration My accreditation organization as well as our internal policies require periodic calibr...

Страница 15: ...pprox What s wrong Nothing When in the Concentrator Mode the electronic components of the indicator automatically compensates for the argon found in the gas produced by an oxygen concentrator As a res...

Страница 16: ...nd continue to hold down pure O2 calibrate button to temporarily remove argon correction from unit s With unit still connected to source of U S P oxygen Display reads between 98 to 102 but not 99 9 on...

Страница 17: ...n Manual Display shows err during use Unit has made a reading outside normal ranges Check calibration Make sure gas being measured is from oxygen concentrator LO BATT appears in display screen when un...

Страница 18: ...be responsible for all shipping charges to return the unit to Salter Labs Salter Labs will be responsible for normal shipping charges for the return of the unit to purchaser All warranties are voided...

Отзывы: