Time  left  display  when  the  PRE-
PROCESSING SETTINGS are disabled 

 
The  time  displayed  in  TIME  LEFT  for  the  Loading 
Station indicates the waiting time required until the 
basket can be transferred to Station 1 after setting 
a  basket  into  the  Loading  Station  (in  other  words, 
the  detention  time  of  the  basket  in  the  Loading 
Station).  This  time  will  vary  depending  on  the 
condition of the preceding basket.  

 

In the image above, if a basket is placed into the Loading Station, 

it will take 23 seconds until the basket will be ready to be 

transferred from the Loading Station to Station #1. 

 

New Workflow 
 

Working  with  the  Pre-Processing  Timer  Software 
will  need  adaption  of  the  workflow  within  the 
laboratory.  When  continuous  loading  of  basket’s  is 
done  the  unloading  will  become  more  important 
while  the  system  will  not  start  with  a  new  basket 
when  four  baskets  are  occupying  the  retorts. 
Therefor  unloading  the  Tissue-Tek

®

  Xpress

®

 

x

120 

(when  occupied  with  four  baskets)  needs  to  be 
done  before  a  new  basket  can  be  loaded  on  the 
Tissue-Tek

®

  Xpress

®

 

x

120.  Make  sure  excessive 

formalin  is  drained  prior  to  loading  the  basket  on 
the Tissue-Tek

®

 Xpress

®

 

x

120. 

Steps in using the PPS timer software 
 

Log on to the Tissue-Tek

®

 Xpress

®

 

x

120 

Fill the loading station with Pre-Processing Solution 

(Item code 7115) 

Press replaced on the touchscreen 
Load basket in the loading station, drain excessive 

formalin  prior  to  loading,  preferably  rinse  with 
Pre-Processing Solution or Alcohol 70%. 

Reload  with  a  new  basket  when  the  Loading  Door 

Icon has turned Green, drain excessive formalin 
prior  to  loading,  preferably  rinse  with  Pre-
Processing Solution or Alcohol 70%. 

When  four  baskets  are  occupying  the  Tissue-Tek

®

 

Xpress

®

 

x

120,  first  unload  a  basket  before 

loading new basket. 

When the Pre-Processing Solution has expired two 

options are available: 
a.

 

When loading station is not occupied with a 
basket, the loading station can be removed 
when  Loading  Door  Icon  is  stated  OK.  Pre-
Processing  Solution  can  be  discarded  to 
local regulations and the loading station can 
be filled with PPS again. When placed back,  
REPLACED  can  be  selected  on  the  touch 
screen message. 

b.

 

When the loading station is occupied with a 
basket, LATER can be selected on the touch 
screen message. When the basket transfers 
to  the  first  microwave  retort  the  previous 
step (7a) can be performed. 

At  the  end  of  the  day  discard  the  Pre-Processing 

Solution to local regulations. 

 

 

 
 
 

Содержание Tissue-Tek Xpress x120

Страница 1: ...on filled with Pre Processing Solution or Molecular Fixative Basket holds up to 40 cassettes Run preparation for processing Pre Start Checks 1 Loading Station is filled with Pre Processing Solution and replaced every 4 hours or filled with Molecular Fixative and refreshed daily depending on the software version 2 No baskets are loaded in the instrument 3 Transfer arm is free for moving 4 All retor...

Страница 2: ...o the holder of the Loading Station Close the door 5 Enter the number of cassettes to be processed in the cassette count data entry dialog display if Reagent Control is set to cassettes 6 Select the sender from the select item menu if user defined file is enabled 7 Place the next basket into Loading Station when empty System Operation Process Monitoring phase 2 A graphical display provides the sta...

Страница 3: ...esh PPS When the Tissue Tek Xpress x120 is fully occupied with four baskets the first basket should be unloaded before a new basket can be loaded Pre Processing Timer Function enabled and ready to load System Operation Unloading Processed Specimen phase 3 1 Verify if the Access Permitted Icon the Unloading Indicator are illuminated and an audible alarm sounds briefly 2 Press the unloading indicato...

Страница 4: ...ge window will not be displayed until the SOLUTION EXPIRATION TIME is met Display during Pre processing in the Loading Station The following statuses are displayed after a basket is loaded in the Loading Station Once a basket is accepted into the Loading Station the Preprocessing TIME LEFT will begin counting down and the robotic arm will conduct the Basket Dipping operation into the preprocessing...

Страница 5: ...sue Tek Xpress x120 This feature is included to prevent tissue specimens ending up in a microwave retort when the samples are not collected from the unloading stations and four baskets with specimens are running Pre processing will not commence until a basket is removed from the unloading station Loading Door Icon Explanation of Icon when the Preprocessing Settings are enabled Green OK icon When t...

Страница 6: ...ake sure excessive formalin is drained prior to loading the basket on the Tissue Tek Xpress x120 Steps in using the PPS timer software Log on to the Tissue Tek Xpress x120 Fill the loading station with Pre Processing Solution Item code 7115 Press replaced on the touchscreen Load basket in the loading station drain excessive formalin prior to loading preferably rinse with Pre Processing Solution or...

Страница 7: ...e Daily 1 Exterior cleaning 2 Retort cleaning 3 Exchange drip tray liner 4 Refill Loading station with Pre Processing Solution when indicated or Molecular Fixative 5 Clean magazines or baskets as described Locking lever replace reagents Replace processing reagent Weekly 40 Working hours or after 1500 cassettes whichever comes first 1 Reagent set change reagent and paraffin need to be replaced as a...

Страница 8: ...ow the block to melt down 2 Remove tissue and place in a cassette identified with the proper accession number 3 It is important to note the thickness of the residual tissue 4 Place the labeled cassette in solvent xylene or xylene substitute for 30 minutes two changes 5 Place the cassette with tissue in a beaker containing absolute alcohol for 30 minutes two changes 6 Remove the cassettes from the ...

Страница 9: ...is required 9 Choose Go to standby for reprocessing 10 If conditions causing the error are recovered the system is ready for reprocessing 11 Select recovery button in utility menu NOTE The PROCESS RECOVERY button is enabled only during the WARMING UP SYSTEM cycle 12 Touch the station 3 and or 4 button as appropriate 13 The process time dialog displays 14 Set time that remained when processing was ...

Страница 10: ...issue that is trimmed to 3 mm thick using the Extended program DO NOT allow the faced tissue blocks to soak in water DO face the block and allow the block to cool on ice DO use plastic lids and remove metal staples and sutures from tissue prior to placement in the cassette DO NOT top up reagents or rotate reagents Reagents to be replaced as a set DO change the reagents and solution after 1500 cass...

Отзывы: