Loading Station Icon & Solution 
Expiration Message 

 
If  the  countdown  reaches  the  SOLUTION 
EXPIRATION 

TIME 

set 

in 

PREPROCESSING 

SETTINGS,  the  message  window  in  the  bottom 
right  screenshot  is  displayed.    (The  above 
statement  is  also  true  even  when  the  unit  is  in 
Stand-by after completing warm-up.)  

The conditions for the Loading Station Icon Colors 
are as follows: 
 

Blue

 – Set Solution Expiration Time has not 

expired.  

Yellow

 – Set Solution Expiration Time has expired 

and needs to be replaced.  

White

 – Solution Expiration Time function is not 

used.   

 

Set Solution Expiration Time has not expired 

 

Set Solution Expiration Time has expired 

 

Items 

Description 

LATER 

Touch LATER to continue without replacing the preprocessing solution.  The 
Loading  Station  icon  color  will  still  be  yellow.    The  message  window  will 
display every 10 mins. (snooze function) until REPLACED is selected. 

REPLACED 

Touch  REPLACED  after  replacing  the  preprocessing  solution.    The  Loading 
Station icon color will change back to blue.   The message window will not 
be displayed until the SOLUTION EXPIRATION TIME is met. 

 

Display  during  Pre-processing  in  the 
Loading Station 

 
The following statuses are displayed after a basket 
is loaded in the Loading Station.  Once a basket is 
accepted 

into 

the 

Loading 

Station, 

the 

Preprocessing  TIME  LEFT  will  begin  counting  down 
and  the  robotic  arm  will  conduct  the  “Basket 
Dipping”  operation  into  the  preprocessing  solution 
in order to reduce the amount of air bubbles in the 
cassettes  and  to  efficiently  mix  the  reagents  with 
the  specimens.  However,  if  there  are  preceding 
baskets  being  transferred,  the  Loading  Station 
“Basket Dipping” operation will not be initiated until 
the transfer of preceding baskets is complete. If the 
basket is removed from the Loading Station during 
Preprocessing,  Preprocessing  is  automatically 
aborted 

 

Counting down the Pre-Processing Time 

 

Items 

Description 

START TIME 

Denotes the time to start transfer to Station 1.  Because the pump out time 
in  Station  1  is  taken  in  to  consideration,  the  time  displayed  is  several 
seconds earlier than the actual start time of the basket transfer.  

TIME LEFT 

Denotes  the  time  remaining  before  starting  transfer  to  Station  1.    It  also 
includes the pump out time required for Station 1.  The time required for 
pump out is included in the set PREPROCESSING TIME.  

 

Содержание Tissue-Tek Xpress x120

Страница 1: ...on filled with Pre Processing Solution or Molecular Fixative Basket holds up to 40 cassettes Run preparation for processing Pre Start Checks 1 Loading Station is filled with Pre Processing Solution and replaced every 4 hours or filled with Molecular Fixative and refreshed daily depending on the software version 2 No baskets are loaded in the instrument 3 Transfer arm is free for moving 4 All retor...

Страница 2: ...o the holder of the Loading Station Close the door 5 Enter the number of cassettes to be processed in the cassette count data entry dialog display if Reagent Control is set to cassettes 6 Select the sender from the select item menu if user defined file is enabled 7 Place the next basket into Loading Station when empty System Operation Process Monitoring phase 2 A graphical display provides the sta...

Страница 3: ...esh PPS When the Tissue Tek Xpress x120 is fully occupied with four baskets the first basket should be unloaded before a new basket can be loaded Pre Processing Timer Function enabled and ready to load System Operation Unloading Processed Specimen phase 3 1 Verify if the Access Permitted Icon the Unloading Indicator are illuminated and an audible alarm sounds briefly 2 Press the unloading indicato...

Страница 4: ...ge window will not be displayed until the SOLUTION EXPIRATION TIME is met Display during Pre processing in the Loading Station The following statuses are displayed after a basket is loaded in the Loading Station Once a basket is accepted into the Loading Station the Preprocessing TIME LEFT will begin counting down and the robotic arm will conduct the Basket Dipping operation into the preprocessing...

Страница 5: ...sue Tek Xpress x120 This feature is included to prevent tissue specimens ending up in a microwave retort when the samples are not collected from the unloading stations and four baskets with specimens are running Pre processing will not commence until a basket is removed from the unloading station Loading Door Icon Explanation of Icon when the Preprocessing Settings are enabled Green OK icon When t...

Страница 6: ...ake sure excessive formalin is drained prior to loading the basket on the Tissue Tek Xpress x120 Steps in using the PPS timer software Log on to the Tissue Tek Xpress x120 Fill the loading station with Pre Processing Solution Item code 7115 Press replaced on the touchscreen Load basket in the loading station drain excessive formalin prior to loading preferably rinse with Pre Processing Solution or...

Страница 7: ...e Daily 1 Exterior cleaning 2 Retort cleaning 3 Exchange drip tray liner 4 Refill Loading station with Pre Processing Solution when indicated or Molecular Fixative 5 Clean magazines or baskets as described Locking lever replace reagents Replace processing reagent Weekly 40 Working hours or after 1500 cassettes whichever comes first 1 Reagent set change reagent and paraffin need to be replaced as a...

Страница 8: ...ow the block to melt down 2 Remove tissue and place in a cassette identified with the proper accession number 3 It is important to note the thickness of the residual tissue 4 Place the labeled cassette in solvent xylene or xylene substitute for 30 minutes two changes 5 Place the cassette with tissue in a beaker containing absolute alcohol for 30 minutes two changes 6 Remove the cassettes from the ...

Страница 9: ...is required 9 Choose Go to standby for reprocessing 10 If conditions causing the error are recovered the system is ready for reprocessing 11 Select recovery button in utility menu NOTE The PROCESS RECOVERY button is enabled only during the WARMING UP SYSTEM cycle 12 Touch the station 3 and or 4 button as appropriate 13 The process time dialog displays 14 Set time that remained when processing was ...

Страница 10: ...issue that is trimmed to 3 mm thick using the Extended program DO NOT allow the faced tissue blocks to soak in water DO face the block and allow the block to cool on ice DO use plastic lids and remove metal staples and sutures from tissue prior to placement in the cassette DO NOT top up reagents or rotate reagents Reagents to be replaced as a set DO change the reagents and solution after 1500 cass...

Отзывы: