background image

BarbLock® Tabletop Pneumatic Assembly Tool: Operations Manual

www.biopharm.saint-gobain.com

15

FLS-5278C-CLWR-Rev 01

20.0 Disassembly

Disconnect the air supply before performing any work on the tool. Lay the tool flat on the left side (see image 19 be-
low). Use a Phillips screwdriver to remove the 8 screws that are visible in the right side Case. Lift the right side Case 
off the tool, along with its screws. Figure 19 shows the appearance of the tool after removal of the right side Case.

Do not insert any hard or sharp tools or other objects into the interior spaces of the BarbLock® Assembly Tool. Doing 
so could damage internal components and cause the tool to malfunction.

Figure 19: Inside of the tool after removal of the right side Case and screws

Figure 20: A close up of view of the grease fitting on end of Carriage

Содержание BarbLock BLT-HHPAD

Страница 1: ...Saint Gobain BarbLock Assembly Tool Part Number BLT HHPAD OPERATIONS MANUAL www biopharm saint gobain com...

Страница 2: ...Assembly Tool 9 Air Supply Connection 11 Operational and Safety Checks 12 Noise Emissions 12 Vibration 12 Tool Balance Connection and Use 12 Residual Risks 13 Use and Storage Conditions 13 Cleaning 1...

Страница 3: ...sole risk Proper use also includes compliance with operating inspection and maintenance conditions prescribed by the manufacturer Non com pliance with the warnings in the operating instructions and i...

Страница 4: ...nnec tion This creates an ultra secure connection that is more resistant to leakage than any other available technology Figure 2 below shows some of the Saint Gobain products with which the BarbLock A...

Страница 5: ...Saint Gobain brand name and symbol Figure 3 2 Product identification and company information Figure 4 3 This User Operator manual contains critical information that must be understood by the user in o...

Страница 6: ...rved the tool should be disconnected from the air supply and removed from the work area When the jaws are in the closed position a pinch risk is present in gaps just below the lower jaw see Figure 11...

Страница 7: ...will move any time the Safety Cover is closed and the trigger is pressed Keep fingers away from all openings in the area of the jaws Failure to do so could result in a serious pinching injury to the f...

Страница 8: ...not present in this image Jaw tail Figure 14 Example of Upper Jaw showing part number with suffix letter A to indicate upper position Jaws are installed in their respective carriers A suffix on the J...

Страница 9: ...ed to the tool This will happen no matter if the Safety Cover is closed or not Be aware of the possibility that the lower jaw carrier may move while connecting the tool to the pressurized air supply K...

Страница 10: ...o need for the operator to provide intervention to the BarbLock assembly WARNING Pinch Hazard after the BarbLock has fully closed the operator can release the Trigger This will cause the jaws to retra...

Страница 11: ...smaller sizes of BarbLocks as long as full closure of the BarbLocks is consistently achieved Pressure should be adjusted as needed to achieve full closure of the BarbLocks in use at the time The User...

Страница 12: ...nting motion of the jaws any time the Safety Cover is open If the Safety Cover interlock switch is defeated then the jaws can move while the Safety Cover is open and a pinch injury becomes more likely...

Страница 13: ...tal limits are applicable to the BarbLock Assembly Tool Use Conditions Storage Transport Conditions Temperature C 0 to 40 10 to 50 Altitude 0 to 6560 2000m 0 to 40 000 12 200m Relative Humidity 0 to 8...

Страница 14: ...35252AL BLTJAW005 PF135312PPAF AL BLTJAW006 BL135374 BLTJAW007 PF135375PPAF BLTJAW007 BL135375 BLTJAW007 BL135375AL BLTJAW007 BL135376 BLTJAW007 BL135377ALLRRPVDF BLTJAW007 BL135500 BLTJAW008 BL135500...

Страница 15: ...are visible in the right side Case Lift the right side Case off the tool along with its screws Figure 19 shows the appearance of the tool after removal of the right side Case Do not insert any hard or...

Страница 16: ...h the kit see figure 22 below Grease is injected into the carriage by inserting the tip of the type L fitting into the open end of the nipple on the end of the carriage shown in the red circle in figu...

Страница 17: ...the joint between the left and right side Cases to confirm that they are properly aligned with each other and the two Case halves are in contact Once this is confirmed insert the 8 screws that attach...

Страница 18: ...y to return to service 24 0 Troubleshooting If the Safety Cover the Latch or the jaws did not function as they should as described in the Functional Confir mation section or if an air leak was detecte...

Страница 19: ...ool Operations Manual www biopharm saint gobain com 19 FLS 5278C CLWR Rev 01 25 0 Tool Physical Details Tool Weight 5 2 pounds including one set of jaws Tool Dimensions refer to image below Figure 23...

Страница 20: ...BarbLock Assembly Tool Operations Manual www biopharm saint gobain com 20 FLS 5278C CLWR Rev 01 26 0 CE Mark Declaration Of Conformity...

Отзывы: