INSTRUCTION MANUAL
Sahara Moisture Transmitter
Reference 085
Rev 2
Page 9 of 14
06/06/2013
VI. Return
Policy
If returning an SMT for warranty service or annual recalibration, the user must contact
Henderson Engineering Co. for instructions, which will include the issue of a Return
Merchandise Authorization (RMA), and the user will issue a Purchase Order for the
itemized service and/or parts. No credit will be issued for items returned without
proper authorization.
Henderson Engineering Co. will inspect all items returned prepaid. If inspection proves
the equipment to be defective, full credit will be issued. If inspection proves the
equipment to be in good working condition, or it is found to be damaged by abuse, an
estimate of the charges will be made before work begins, provided the equipment is
not covered under warranty. Upon completion of work as approved by user, the
equipment will be returned F.O.B. factory, and payment for items issued against the
user's purchase order will be required.
VII. Helpful
Information
A. Ten Things to Consider When Making Moisture Measurements in Gases
1. The sampling system should be allowed to equilibrate for up to (typically) 24
hours.
2. A purge flow meter aids in determining equilibration point and will indicate
integrity of seals. If probe reading changes with a change of flow rate:
a.
Equilibration has not been achieved.
b.
System is leaking.
c.
Both of above.
3. Measurements should be made at system pressure, or highest possible
pressure, to minimize outgassing of metal parts, wall effects, etc. This will
result in higher dew points with increased sensitivity. Remember, it’s
necessary to convert readings taken at atmosphere to equivalent dew point at
line pressure, if line pressure dew point is what you want, or vice versa. For
example, -40
F at 100 psig line pressure is equal to -70
F at atmospheric
pressure.
Содержание Sahara-Pak HC-1260
Страница 40: ...INSTRUCTION MANUAL Model HC Reference 219 Rev 2 Page 40 of 46 06 20 2013 SECTION 4 WARRANTY...
Страница 48: ...63116 Airdyne Inc 14910 Henry Rd Houston TX 77060 Alcon Labs 9965 Buffalo Speedway Houston TX 77054...
Страница 49: ...63116 Airdyne Inc 14910 Henry Rd Houston TX 77060 Alcon Labs 9965 Buffalo Speedway Houston TX 77054...
Страница 50: ......
Страница 51: ...63116 Airdyne Inc 14910 Henry Rd Houston TX 77060 Alcon Labs 9965 Buffalo Speedway Houston TX 77054...
Страница 52: ...63116 Airdyne Inc 14910 Henry Rd Houston TX 77060 Alcon Labs 9965 Buffalo Speedway Houston TX 77054...
Страница 53: ...63116 Airdyne Inc 14910 Henry Rd Houston TX 77060 Alcon Labs 9965 Buffalo Speedway Houston TX 77054...
Страница 54: ...63116 Airdyne Inc 14910 Henry Rd Houston TX 77060 Alcon Labs 9965 Buffalo Speedway Houston TX 77054...
Страница 87: ...INSTRUCTION MANUAL Model HC Reference 219 Rev 2 Page 45 of 46 06 20 2013 SECTION 7 CUT SHEETS...
Страница 128: ...INSTRUCTION MANUAL Model HC Reference 219 Rev 2 Page 46 of 46 06 20 2013 SECTION 8 DRAWINGS...
Страница 129: ......
Страница 130: ......
Страница 131: ......
Страница 132: ......
Страница 133: ......
Страница 134: ......
Страница 135: ......
Страница 136: ......