PIDLOGGER - PID OPTIMIZATION

(for skilled personnel only)

The PID parameters are factory optimized for most of use. However it's possible to change them to best suite particular
conditions of:
- heavy load (heavier load means greater thermal inertia)
- fresh air on or off

To change the PID parameters hod down together

and

button for 6 seconds.

Insert the following parameter: proportional band at 40°C ("1 ??"), proportional band at 280°C ("2 ??"), integral time ("3 ???"),
derivative time("4 ???")
Set parameter values by pressing

to select digit and

to modify the value

By pressing

the parameters will be memorized otherwise if you hold down

for 2 seconds you will

exit from the parameter setting function.

Example of PID parameters

"1 10"

(proportional band at 40°C is +-10°C)

"2 05"

(proportional band at 280°C is +-5°C)

"3 150"

(integral time is 150 seconds)

"4 37"

(derivative time is 37 seconds)

The proportional band (BP) is calculated with linear interpolation. For example at 100°C it is:

BP@100°C = BP@40°C + (BP@280°C - BP@40°C) * (100 - 40) / (280 - 40) - 40) / (280 - 40)

DRYCONTROL - ON-OFF OPTIMIZATION

(for skilled personnel only)

To change the PID parameters hod down together

and

button for 6 seconds.

Insert the following parameter: setpoint_low (on the left) and setpoint_high(on the right)
?.?-?.?
Set parameter values by pressing

to select digit and

to modify the value

The parameter will be accepted if setpoint_high > (setpoi 0.5%RH)
By pressing

the parameters will be memorized otherwise if you hold down

for 2 seconds you will

exit from the parameter setting function.

The factory parameter of the ON-OFF humidity control are:
- setpoint_low = 4.5%RH
- setpoint_high = 5.0%RH

ERROR MESSAGES

The PIDLogger and Drycontrol warn when one of the following error is happened:

Err 1:

230Vac is not detected at the heating elements when the PID wants to heat the oven. Possible cause is an interruption of

the Solid State Relay (SSR)

Err 2:

230Vac is detected at the heating elements when the PID doesn't want to heat the oven. Possible cause is a short circuit

of the Solid State Relay (SSR). It's a dangerous condition because the oven heat without control and only the Alarm Set Point
can stop him.

Err 5:

Humidity control activated but the RH sensor is not inserted

Sahara DRY, 07/.2021

Page 13

Содержание 80VF

Страница 1: ...SAHARA DRY 80VF 120VF 250VF FORCED VENTILATION BAKING OVEN with HUMIDITY CONTROL from 80 to 250 litre 40 280 C Instruction Manual Sahara DRY English...

Страница 2: ...onform to EC directives directives EMC 2014 30 CE LVD 2014 35 CE MD 2006 42 CE the following harmonized standards are in use EN 60204 1 EN 61000 6 1 EN 61000 6 3 EN 292 A weighted sound pressure 70dB...

Страница 3: ...372 477 522 Height mm 472 512 797 External Dimensions Width mm 796 836 956 Depth mm 570 680 730 Height mm 680 720 1025 Temperature Control P I D P I D P I D Humidity control ON OFF at T 100 C ON OFF...

Страница 4: ...ing function to change selected digit to modify value to confirm LED Air shows when the eletrovalve is ON and the dry air is insufflated LED shows the condition when ON OFF control is working relative...

Страница 5: ...rnal chamber when you want to reach temperatures of over 100 C FRESH AIR EXCHANGER Setting of air exchanger Air output 35mm diameter is placed on the rear side of the oven The opening and closing of t...

Страница 6: ...ressing to select digit and to modify the value Bypressing theparameterwillbememorized otherwiseifyouholddown for2secondsyouwillexit from the parameter setting function Example of setpoint at 195 C t...

Страница 7: ...you will exit from the parameter setting function At the end the display will show START TIME Press and insert the following parameters start month M start day d start hour h start minutes m Set param...

Страница 8: ...switched off in two ways 1 usingthemainswitchON OFF AtthenextswitchontheovenwillbeagaininProgmodeandthePIDwillworkimmediately 2 using the STOP mode Hold down for 2 seconds the display will show Prog P...

Страница 9: ...mpressed air The Eco mode is provided for a more rational use of compressed air in fact the low humidity is reached for cycles later and not in a direct way Deactivate the ON OFF humidity control Hold...

Страница 10: ...er exe inadirectoryofyourPCandrun it to be operative SETTING THE PIDLOGGER CONNECTION 1 Connect the Sahara oveb to the PC using the specific USB cable 2 Set the oven in Com On mode 3 Run PIDManager pu...

Страница 11: ...e end of memory is reached 21600 samples Every tyme SETPOINT COUNTDOWN and PROG mode are activated all previous data in memory are erased and overwritten The sample data are written inside non volatil...

Страница 12: ...button export to one of the 10 program memorized into the Sahara oven with Export button In every step of a program is possible to execute one of these 3 commands SET POINT the oven reach a setpoint t...

Страница 13: ...RYCONTROL ON OFF OPTIMIZATION for skilled personnel only To change the PID parameters hod down together and button for 6 seconds Insert the following parameter setpoint_low on the left and setpoint_hi...

Страница 14: ...all parts have reached a normal temperature Use a detergent with less than 5 of non ionic surfactant Lubricate the door inge and the closing system of the handle SPARE PARTS 8107 202 Pidlogger 8107 2...

Страница 15: ...PIDLogger DRYControl CONNECTIONS SaharaDRY 07 2021 Page15...

Страница 16: ...only Page16 SaharaDRY 07 2021 P I D control board EMI Filter Mains Input Fuses SSR 12V Yellow Lamp Alarm Reset Button Red Lamp Alarm Relay R R 1 R 2 Heating Resistance Fan Motor Transformer Over Tempe...

Отзывы: