The sample duration depends on the sample rate.
Sample rate
Duration
(soloT)
Duration (T + RH%)
1 sec
6 hours
3 hours
5 sec
1 day e 6 hours
15 hours
10 sec
2 days e 12 hours
1 day and 6 hours
15 sec
3 days e 18 hours
2 days and 12 hours
30 sec
7 days e 12 hours
3 days and 18 hours
1 min
15 days
7 days and 12 hours
5 min
75 days
37,5 days
10 min
150 days
75 days
15 min
225 days
112,5 days
30 min
450 days
225 days
ATTENTION
: when you set the new value of date, time and sample rate any collected data logged into the PIDLogger memory
will be erased.
DATALOGGER SAMPLE STARTING
The Sahara oven starts to sample when one of these mode is activated:
- SETPOINT
- COUNT-DOWN
- PROG
The sampling activity is interrupted when STOP mode is activated or the end of memory is reached (21600 samples).
Every tyme SETPOINT, COUNTDOWN and PROG mode are activated all previous data in memory are erased and
overwritten.
The sample data are written inside non-volatile memory so are readable even after a black-out.
DATA DOWNLOAD
Pull down the "Datalogger" menu and click on "Read" for downloading and displaying the collected temperature data.
Immediately after the downloading a message prompt will ask you if you want to save the data.
The format used to save the data is "*rec" that is a simply text file like this:
***TEMP,TIME***
Interval=5 Seconds
158.1,2010/10/04 17:11:25
158.5,2010/10/04 17:11:30
158.9,2010/10/04 17:11:35
....
This file can be easily edited into a text editor or imported into a spreadsheet.
PRINTING
To print the showed graph pull down the "File" menu and click on "Print".
OPENING OF A FILE
Pull down the menu "File", click on "Open" and insert the path of the "*.rec" file to open.
Sahara DRY, 07/.2021
Page 11