Display
Meaning
Action
Low Temperature
Display
High Ambient
Temperature
Display
Low Ambient
Temperature
Display
Error Function
Display
Low Battery
Display
No display or
fixed display.
Measured temperature too
low, 3 short beeps are emitted.
Ambient temperature too high
(above 104.0˚F or 40.0˚C). 3
short beeps are emitted.
Ambient temperature too low
(below 50.0˚F or 10.0˚C). 3
short beeps are emitted.
Thermometer has
malfunctioned. 3 short beeps
are emitted.
Battery voltage is low.
Temperature measurement is
still allowed.
The thermometer should be
reset or the battery should be
replaced.
Probe may have been inserted
incorrectly. Retake temperature
measurement.
Take the measurement in
operating temperature range.
(50.0˚F -104.0˚F / 10.0˚C -
40.0˚C)
Take the measurement in operating
temperature range. (50.0˚F -
104.0˚F/ 10.0˚C - 40.0˚C)
Turn the thermometer off and
re-take temperature measurement.
Replace old battery with new
battery after taking the
measurement.
Replace the battery or reset the
thermometer by turning the
thermometer off and on again.
Troubleshooting
Temperature measurement range
89.6°F – 108.0°F (32.0°C – 42.2°C)
Operating temperature range
50.0˚F - 104.0˚F (10.0˚C - 40.0˚C)
Ear Thermometer meets accuracy requirements established in ASTM Standards E1965-98.
Batteries:
one 3V lithium battery, CR 2032.
A clinical summary is available upon request. For home use only.
Storage temperature range: -4.0˚F - 122.0˚F (-20˚C - 50˚C)
ASTM laboratory accuracy requirements in the
display range of 98.6°F to 102.2°F (36°C to
39°C) for infrared thermometers is ±0.4°F
(±0.2°C), whereas for mercury-in-glass and
electronic thermometers, the requirement per
ASTM Standards E667-86 and E1112-86 is
±0.2°F (±0.1°C).
Technical Specifications
Maximum Laboratory Error
Temperature
Maximum Error
Fahrenheit Scale:
96.8˚F to 102.2°F
±0.4°F
Outside this range
±0.5°F
Celsius Scale:
36˚C to 39°C
±0.2°C
Outside this range
±0.3°C
3