Connect the incubator to an earthed power supply using the power cord provided. The display 
will illuminate immediately. NB:- Do not programme the CO

2

 level yet 

 

 
3.10 

The chamber set points are pre-programmed at 37.0°C and 0% CO

2

. Leave the incubator 

until the programmed chamber temperature has been reached. 

 
3.11 

Carefully  unpack  the  humidity  tray.  Fill  with  0.5  litres  of  warm  distilled  water  (at  approx. 
37.0°C).    For  cell  culture  work  –  we  recommend  the  use  of  copper  sulphate  (or  recognised 
biocide) in the humidity tray (one small teaspoonful copper sulphate recommended). 
For IVF and other sensitive work – we do not recommend the use of any biocide in the 
humidity tray. To reduce the possibility of contamination, the humidity tray water should be 
changed and the humidity tray cleaned with 70% Alcohol / 30% distilled water solution every 
10 – 14 days. To remove / replace the humidity tray: - turn the tray through 45

°

 and remove 

for refilling.. To fit the humidity tray - Turn the humidity tray through 45

°

 and pull it forward as 

shown in the pictures on the following page:- 
 

 

 

 

NB – ensure that the Humidity Tray is located centrally between the racks and pushed fully 
back against the stop as shown above right. 

 
3.13

 

Fit the shelves and leave the incubator for at least two hours (preferably overnight) to allow 
conditions to stabilise.  

 
3.14 

Programme the CO

2

 level (i.e. 5%) and ensure that the CO

2

 level stabilises at the 

programmed level 

 

3.15 

To carry out the Auto Zero, press the 

 and 

 keys simultaneously. The Inner Door must 

be opened and closed according to the displayed instructions, after which the process is 
completely automatic. 

 
3.16 

The incubator is now ready for use. 

 

Содержание Mini Galaxy E 050-600

Страница 1: ...CO2 Incubator Model No 050 600 User Manual...

Страница 2: ...5 Control Panel Description 8 6 Programming Temperature CO2 Auto Zeroing 9 7 Programming the Alarm System 10 8 The Alarm System 11 9 System Alarms 13 10 Cleaning and Disinfecting 14 11 Technical Spec...

Страница 3: ...o control the level of CO2 giving excellent reliability and is unaffected by humidity The CO2 system has a semi automatic zero function which provides a simple process for the user to follow to mainta...

Страница 4: ...midity tray cleaned with 70 Alcohol 30 distilled water solution every 10 14 days To remove replace the humidity tray turn the tray through 45 and remove for refilling To replace the humidity tray care...

Страница 5: ...t are supplied fitted to the unit 4 Anti slip pads for the adjustable feet 1 Black sensor cover in situ If required further shelves can be purchased up to a maximum of 6 The humidity tray has a fixed...

Страница 6: ...ully remove any packaging Fit 7 silicone rubber suction cup feet to the shelf racks in positions indicated on the picture below left with an Repeat for the opposite shelf rack To fit the shelf racks t...

Страница 7: ...ray water should be changed and the humidity tray cleaned with 70 Alcohol 30 distilled water solution every 10 14 days To remove replace the humidity tray turn the tray through 45 and remove for refil...

Страница 8: ...growth in the tray this can reduce contamination on the chamber walls one small teaspoonful copper sulphate recommended For IVF and other sensitive work we do not recommend the use of any biocide in...

Страница 9: ...3 segment LED displays and 4 keys The functions and descriptions of these are described below PROGRAMMING Key PLUS Key MINUS Key ACCEPT Key The decimal points on the display for Temperature and CO2 f...

Страница 10: ...ft causing an inaccuracy in the CO2 level at the programmed value It is recommended that the CO2 system is Auto Zeroed monthly to ensure that CO2 level is correct To carry out the Auto Zero press the...

Страница 11: ...the display will show doo doo doo doo rAL rAL rAL rAL Press the Key and the Keys to adjust the time in the following steps 15 30 45 60 75 90 seconds then OFF Press to select the preferred value Press...

Страница 12: ...pened the Alarm System is disabled and on closing the door the preset Alarm Arm timeout starts If chamber conditions recover within the Alarm Arm Time the Alarm System is rearmed At the timeout the Al...

Страница 13: ...SE INNER DOOR 15 MINUTES ALARM DELAY TIMEOUT STARTED IS TEMPERATURE OR CO2 LEVEL MORE THAN 0 5 OUTSIDE PROGRAMMED POINTS CANCEL ALARM BY PRESSING ANY KEY ALARMS CANCELLED VISUAL AND AUDIBLE ALARMS CAN...

Страница 14: ...ntrol valve The condition is shown on the display by the message Ct Ct Ct Ct riP riP riP riP When the chamber temperature has fallen to the programmed set point the system changes to mode 2 Mode 2 The...

Страница 15: ...can be cleaned by rinsing in sterile water wiping down with the alcohol solution and rinsing with sterile water 10 9 Wipe down the inside of the chamber with alcohol water solution and leave to dry c...

Страница 16: ...re to 50 C Control 0 1 C Stability 0 1 C Uniformity 0 2 C CO2 Control Solid state infra red CO2 sensor operating independent of humidity Programmable fully automatic zeroing facility Range 0 2 10 or 0...

Страница 17: ...kg Electrical Supply Earthed Electrical Supply of Voltage 220 240v 50 60Hz Or 100 120v 50 60Hz Power 500 watts Energy to maintain 37 C 0 1kwh Storage Temperature 10 50 C NB Factory calibration of the...

Страница 18: ...cubator 0 2 10 CO2 220 240V 50 60 Hz Model 050 600 100 120V 50 60 Hz Model 050 601 No Charge Factory Fitted Options CO2 Control 0 2 20 050 513 Left Hand Door hinged on right 050 527 Model Options Acce...

Страница 19: ...tware RS232 Port 050 117 Stacking Frame for double stacking 050 118 Multi position Shelf Racks Max 6 positions 050 121 Automatic CO2 Cylinder Change over Unit 170 101 CO2 Two stage Regulator 170 102 C...

Отзывы: