16 

 
Alarm Systems 

 
Two-level alarm system giving programmable audio / visual warnings with options for remote 
communication. Level 1 signals system failures, level 2 is programmable and monitors chamber 
conditions. 

 

Dimensions 

 
Seamless fully radiused interior with removable door seal  
 

 

 

Chamber (HxWxD): 

 

401 x 401 x 308mm 

 
Chamber Volume: 

 

48 litres 

 
External (HxWxD): 

 

645 x 484 x 470mm 

 
Crated Dimensions (HxWxD): 

790 x 620 x 610mm (Including pallet)  

 
Weight (uncrated): 

 

32kg (approx – actual weight will depend on options fitted).  

 
Weight (crated):  

 

50kg 

 

Electrical Supply 

 
Earthed Electrical Supply of: 
Voltage: 

 

 

220 - 240v 50/60Hz 

 

Or 100 - 120v 50/60Hz   
 

Power:   

 

 

500 watts  

 

 

 

 

 

 
Energy to maintain 37°C: 

< 0.1kwh. 

 

Storage Temperature 

 
 

 

 

 

10 - 50°C 

 
 
 
 

NB:- 

Factory calibration of the incubator is carried out at 37°C, 5% CO2 & 95% RH in an ambient 

temperature of 20 - 25°C, with no heat-generating apparatus inside the chamber. 
 
Software calibration adjustments may be required to optimise performance if the incubator is being 
used well outside of these operating conditions. Performance specifications may also be affected. 
 
Please contact RS Biotech, quoting the incubator Serial Number and complete details of operating 
conditions, for advice on calibration adjustments and relevant performance specifications. 
 

Содержание Mini Galaxy E 050-600

Страница 1: ...CO2 Incubator Model No 050 600 User Manual...

Страница 2: ...5 Control Panel Description 8 6 Programming Temperature CO2 Auto Zeroing 9 7 Programming the Alarm System 10 8 The Alarm System 11 9 System Alarms 13 10 Cleaning and Disinfecting 14 11 Technical Spec...

Страница 3: ...o control the level of CO2 giving excellent reliability and is unaffected by humidity The CO2 system has a semi automatic zero function which provides a simple process for the user to follow to mainta...

Страница 4: ...midity tray cleaned with 70 Alcohol 30 distilled water solution every 10 14 days To remove replace the humidity tray turn the tray through 45 and remove for refilling To replace the humidity tray care...

Страница 5: ...t are supplied fitted to the unit 4 Anti slip pads for the adjustable feet 1 Black sensor cover in situ If required further shelves can be purchased up to a maximum of 6 The humidity tray has a fixed...

Страница 6: ...ully remove any packaging Fit 7 silicone rubber suction cup feet to the shelf racks in positions indicated on the picture below left with an Repeat for the opposite shelf rack To fit the shelf racks t...

Страница 7: ...ray water should be changed and the humidity tray cleaned with 70 Alcohol 30 distilled water solution every 10 14 days To remove replace the humidity tray turn the tray through 45 and remove for refil...

Страница 8: ...growth in the tray this can reduce contamination on the chamber walls one small teaspoonful copper sulphate recommended For IVF and other sensitive work we do not recommend the use of any biocide in...

Страница 9: ...3 segment LED displays and 4 keys The functions and descriptions of these are described below PROGRAMMING Key PLUS Key MINUS Key ACCEPT Key The decimal points on the display for Temperature and CO2 f...

Страница 10: ...ft causing an inaccuracy in the CO2 level at the programmed value It is recommended that the CO2 system is Auto Zeroed monthly to ensure that CO2 level is correct To carry out the Auto Zero press the...

Страница 11: ...the display will show doo doo doo doo rAL rAL rAL rAL Press the Key and the Keys to adjust the time in the following steps 15 30 45 60 75 90 seconds then OFF Press to select the preferred value Press...

Страница 12: ...pened the Alarm System is disabled and on closing the door the preset Alarm Arm timeout starts If chamber conditions recover within the Alarm Arm Time the Alarm System is rearmed At the timeout the Al...

Страница 13: ...SE INNER DOOR 15 MINUTES ALARM DELAY TIMEOUT STARTED IS TEMPERATURE OR CO2 LEVEL MORE THAN 0 5 OUTSIDE PROGRAMMED POINTS CANCEL ALARM BY PRESSING ANY KEY ALARMS CANCELLED VISUAL AND AUDIBLE ALARMS CAN...

Страница 14: ...ntrol valve The condition is shown on the display by the message Ct Ct Ct Ct riP riP riP riP When the chamber temperature has fallen to the programmed set point the system changes to mode 2 Mode 2 The...

Страница 15: ...can be cleaned by rinsing in sterile water wiping down with the alcohol solution and rinsing with sterile water 10 9 Wipe down the inside of the chamber with alcohol water solution and leave to dry c...

Страница 16: ...re to 50 C Control 0 1 C Stability 0 1 C Uniformity 0 2 C CO2 Control Solid state infra red CO2 sensor operating independent of humidity Programmable fully automatic zeroing facility Range 0 2 10 or 0...

Страница 17: ...kg Electrical Supply Earthed Electrical Supply of Voltage 220 240v 50 60Hz Or 100 120v 50 60Hz Power 500 watts Energy to maintain 37 C 0 1kwh Storage Temperature 10 50 C NB Factory calibration of the...

Страница 18: ...cubator 0 2 10 CO2 220 240V 50 60 Hz Model 050 600 100 120V 50 60 Hz Model 050 601 No Charge Factory Fitted Options CO2 Control 0 2 20 050 513 Left Hand Door hinged on right 050 527 Model Options Acce...

Страница 19: ...tware RS232 Port 050 117 Stacking Frame for double stacking 050 118 Multi position Shelf Racks Max 6 positions 050 121 Automatic CO2 Cylinder Change over Unit 170 101 CO2 Two stage Regulator 170 102 C...