18
CH1/CH2 Series Mode
Instruction
:
In the series mode, CH1 and CH2 are linked internally
into one channel controlled by CH1. The output
voltage value is twice compared with that of single
channel.
Output ratings
0~60V/0~3A (Max: 64V, 3.2A)
Operation steps
:
1. Press the SER to start the Series mode, and the indicator light
turns bright
2. Connect the load to the positive terminal of CH2 and the
negative terminal of CH1
3. Turn on CH1 and set its current value to 3A. In the default setup,
the instrument works in
“Coarse”, you can turn to “Fine” by
pressing FINE
Coarse: 0.1V or 0.1A each step
Fine: the least precision each step
4. Turn on CH1and set the output voltage/current value by
revolving the
multi-function knob
5. Press ON/OFF button to open the output.
Note:
You can identify the current working state
“CC” or “CV” by
referring to the indicator light.(red means CV, green means CC).