Documenting Multifunction Calibrator / EN
26
06/04/2018 Version No. 1
8a-2. Measure and Output Loop Power 24V
(1) Turn on the power. Turn sliding switch to mA.
(2) Press
button to select MEASURE (input) mode.
(Press
to set the OUTPUT mode as SOURCE,
as for HART
TM
mode users can set it as Enabled or Disabled.)
(3) Connect Test leads or Alligator clips to below terminals:
Red to
“mA V Hz +LOOP” terminal; Black to “mA V” terminal.
(4) Then connect the other ends of Test leads or Alligator clips to the object
for measurement.
(5) The calibrator display will show the measurement result.
(6) Data logging function: refer to
“Data logging” chapter.