Documenting Multifunction Calibrator / EN
18
06/04/2018 Version No. 1
Remark:
1.
Users are allowed to type in max. 5 digits.
2.
Type in each parameter (including decimal point), press ENTER, then the
calibrator will output the parameter values.
3.
When the output value is <0, please type in a negative sign first.
4.
When the output value is <1 and >0,
please type in “0.” first.
Warning:
1.
Do not make an input or connect any charged installment to SOURCE
(terminals) to prevent from damaging the calibrator.
2.
When there is an output short circuit or overload, the calibrator can not
output the correct frequency.
3.
Perform one function at a time and make connection to the specific
terminals only. Remove all other connections to the unused terminals.
Always connect to only one of SOURCE , MEASURE , or TC alone.
4b. How to set up
(1) Press
to enter SETUP function.
(2) Waveform Index: give a number for a
“user programmable waveform”.
(3) Press
again to enter the main display of Hz range.
(4) Press
button to select the
“user programmable waveform”.
Then, the calibrator will output this selected
“user programmable waveform”
(e.g. number 7 waveform
– named SINONE60 Sinusoidal Wave 60Hz).
Содержание RS-133
Страница 1: ...Instruction Manual RS 133 Documenting Multifunction Calibrator Arbitrary Function Generator ...
Страница 40: ...Documenting Multifunction Calibrator EN 34 06 04 2018 Version No 1 Connecting Leads for Multi step Scanning ...
Страница 44: ...Documenting Multifunction Calibrator EN 38 06 04 2018 Version No 1 Connecting Leads for Multi step Scanning ...
Страница 48: ...Documenting Multifunction Calibrator EN 42 06 04 2018 Version No 1 Connecting Leads for Multi step Scanning ...