DOC: OM_VIAP Vienna Air Pans (12A)
6.
First Steps
6.1.
Control of delivery
Checking for transport damage at the time of delivery is recommended.
Any damages are to be recorded and confirmed in writing at the time of delivery and signified from the
carrier. For any damages arising after unpacking, you are obliged to immediately communicate this in
a written report and a telephone call to the supplier.
Transport is covered under your responsibilities. Failing to report any transport damages with the
currier within the set time leads to the loss of responsibilities by the carrier.
6.2.
Installation
Don’t tilt the refrigerated unit during the installation, the compressor unit could be damaged (only self
contained units). Completely remove the plastic film before start up. Take care that your counter frame
will carry the weight of the equipment and fix the unit. It is essential that the countertop is perfectly
level before the unit is mounted.
Accurate installation is necessary for a trouble-free start-up of the refrigerated unit. Every installation
has to fit to the local electric, safety and hygiene instructions:
Please assemble the unit away from any heat sources or on any places, where you will have
direct sunshine.
Do not place the unit near any equipment which produce a lot of steam (i.e. dishwasher, etc.).
Choose a place which is well ventilated and free of dust.
The room temperature must not exceed 77°F over a long period of time.
The relative humidity must not exceed 60% over a long period of time.
PLEASE ENSURE THAT THE AIR-INTAKE AND OUTLET VENTS (GRILLE) FOR
VENTILATION ARE WIDE ENOUGH AND MUST NEVER BE BLOCKED!
It is recommended that the vents are double the size of the condenser surface!
6.3.
Power plug finished devices
The refrigerated unit is provided with a 6 feet long 3-wire power cord with plug and it may only be
connected to a power socket with a nominal voltage 115 V / single phase / 60 cycles. The metal
construction requires the use of a power socket with safety grounding at all times. The electrical power
supply is to be secured with a 16 ampere circuit breaker.
THE CONNECTION TO ANOTHER VOLTAGE OR FREQUENCY IS NOT
PERMITTED!
THE INSTALLATION OF REMOTE REFRIGERATED UNITS HAS TO TAKE
PLACE VIA A NOMINATED REFRIGERATION SPECIALIST.