Cuff Screw
Connection Methods
Wrist Blood Pressure Monitor:
1. Loosen the screw that secures the cuff to the monitor.
2. Remove the cuff from the monitor.
a. Push cuff to the screw direction.
b. Slightly pull cuff to remove the cuff from the monitor.
3. Connect the tube between the calibrator and the monitor to the air
outlet.
4. Connect the data link cable between the calibrator and the monitors
data link port.
kPa mmHg
Readout
Auto calibration
Pressure gauge
Check
count
Clear
count
Set
count
Operation Procedures
1. Data link status operation
Before operation, make sure the air tube and the data link cable are
securely connected to the calibrator and the blood pressure monitor.
Press the ON/OFF key of the calibrator, then press the M key of the
blood pressure monitor.
Note:
If there is no M key on the blood pressure monitor, please press
the ON/OFF key instead.
A loop will appear on the LCD display of the Blood Pressure Monitor,
indicating the data link status.
Data Link Cable
Air Tube
Date Link Port
Air Tube Socket
Calibrator
Blood Pressure Monitor
Calibrator Screen
Blood Pressure
Monitor Screen
Wrist Cuff
Wrist Cuff